Bayesian statistics integrates prior information in parametric optical modeling

A new approach uses embedded data from reference metrology to reduce parametric correlation and improve measurement performance.
11 August 2009
Richard Silver, Bryan Barnes, Nien Fan Zhang, and Hui Zhou

Semiconductor manufacturing requires measurements to control the size of nanoscale devices (referred to as the critical dimension: CD) within patterned layers and for monitoring the placement of new levels upon existing layers (known as overlay metrology). As device sizes continue to decrease, new approaches are required to refine these measurements. A significant body of recent research has investigated new optical technologies for CD and overlay metrology on scales of 32nm and smaller. Much of this work has focused on scatterometry and, more recently, scatterfield microscopy, a technique combining well-defined angle-resolved illumination with image-forming optics.1,2 These optical methods are of particular interest because of their nondestructive, high-throughput characteristics and their potential for excellent sensitivity and accuracy. However, their measurement uncertainties are fundamentally limited by the underlying cross-correlations between the different fit parameters, e.g., line widths and heights.3

To reduce parametric correlation and improve measurement performance and uncertainties, we have developed a Bayesian statistical approach that integrates a priori information gleaned from other measurements. This allows us to embed information obtained from reference metrology and complimentary ellipsometry of the optical constants, or to constrain the floating parametric range based on physical limits or known manufacturing variability. We have implemented this approach using scatterfield microscopy, but it applies equally to scatterometry or methods such as scanning-electron microscopy.

The scatterfield-microscopy instrument is based on a Köhler illuminated bright-field microscope, such that each point at the conjugate back focal plane maps nominally to a plane wave of illumination at the sample.4 Access to a large conjugate back focal plane enables engineered illumination. This, in turn, has fueled advances in optical-system performance, characterization, and data analysis. As a result, the microscope-illumination and collection-path errors can be mapped to a functional dependence. They can then be used to normalize the experimental data for accurate comparison with electromagnetic simulations.5 We acquired both microscope images and backgrounds as a function of angle. We calculated the mean intensity of the angle-resolved images, which we corrected using the background scan that was previously normalized by the known silicon reflectance. This is similar to conventional scatterometry, except that measurements were made with high-magnification image-forming optics.


Figure 1. Experimental data and library data fits for three die from the overlay-metrology advisory group OMAG-3 wafer. Top and middle critical dimensions (CDs) show good agreement with reference values.

We compared the normalized experimental signatures with electromagnetic scattering simulations using parametric analysis. We assembled a library of curves by simulating a multidimensional parameter space. We completed the comprehensive simulations using a rigorous coupled-waveguide analysis (RCWA) model.6 Historically, researchers have used least-squares fitting routines to select a set of parameters with the closest experiment-to-theory agreement.3 Parametric correlation, measurement noise, and model inaccuracy all lead to measurement uncertainty in the fitting process. Even when a measurement demonstrates good sensitivity, cross-correlation among the parameters can lead to very large uncertainties and is the fundamental limitation to the goodness of fit.

Figure 1 shows example data and fit results. They demonstrate good agreement between the simulated library of curves and the experimental data. The goodness-of-fit values and residuals are acceptable. However, the fitting process produces more uncertainty than desired, 1 to 2nm (1σ). In this case, the combination of parametric correlation between fit parameters and measurement noise introduces uncertainty in the final results.

To examine these correlation effects in more detail, we observed a more complicated 3D 22nm contact-hole simulation example with several layers and optical constants. We used a simulation-based analysis comparing spectroscopic and angle-resolved scatterometry.7 To better understand the effects of parametric correlation, we looked at the sensitivity and covariance matrices from the χ2 analysis. The upper part of Figure 2 shows reflectivity data calculated for spectroscopic scans, while the middle graphs show angle-resolved scatterometry results. The uncertainties and covariance matrix are shown at bottom. The angle scans have large regions of excellent sensitivity, yet wavelength scans have smaller uncertainties. The large cross-terms in the covariance matrix lead to larger angle-scan uncertainty values for CD, sidewall, and over etch.


Figure 2. Transverse-electric (TE) and transverse-magnetic scans for different sidewall angles varied by 0.5°. The separation of the curves in the lower graphs demonstrates that the angle scans are more sensitive to 2nm changes in CD than lambda (wavelength) scans. Relatively high cross-correlations among the parameters in angle scans (shown mathematically in the covariance matrix at bottom) result in smaller uncertainties through lambda scanning instead, despite the apparently poor sensitivity from the graphs at top. λ: Wavelength. σCD, σsidewall, and σoveretch: Uncertainly in CD (in nm), sidewall angle (in °), and depth of the hole, which—if not perfect—over etches into another layer below (in nm).

To decouple the parametric correlation and reduce measurement uncertainties, we developed a strategy that embeds multiple measurement uncertainties directly in the parametric fitting and uncertainty analysis. In general, a complete set of measurements consists of N data points obtained under a varying set of conditions for both simulations and experiments. The K model parameters are expressed as a vector and represent the model input parameters, for example, CD and sidewall. N measured values of Y are denoted as {y1, …, yN} and N simulated values with the ith data point xi correspond to the simulated values . We wanted to compare the measured {y1, …, yN} with simulated and find the optimal estimator of the parameter vector . Using a first-order Taylor expansion, the linear approximation of the nonlinear regression can be expressed as  (1)

where εi(0) is an approximation error that we want to minimize.8 By re-parameterization, the model can be expressed as  (2)

where  (3)

with and .8 Referring to the regression model in equation 1, we can treat prior information on β1 as another ‘data point.’9,10 We then get a new and parameter estimates which now include the a priori information. (The variance-covariance matrix of the experimental values is denoted by .) We then obtain new uncertainties on the parameter estimates from the variance matrix, as before.

Using this knowledge, we applied the Bayesian embedded-metrology approach to the silicon target used in Figure 1. Table 1 shows the best-fit values and uncertainties for the regression analysis, with and without embedded atomic-force-microscopy (AFM) reference metrology.11,12 The data shows a change in the mean values as well as an improvement in the uncertainties. The embedded reference measurement data pulls the optical CD scatterfield results toward the AFM-measured values. The resulting uncertainties from the combined measurements are lower than for each of the individual results.

Table 1. Chart showing optical CD (OCD) measurements with and without embedded atomic-force microscopy (AFM) reference metrology. The data shows reduced uncertainties in the combined analysis for the 300nm pitch targets. σTop, σMiddle, and σBottom: Uncertainty in the top, middle, and bottom.
  OCD FittingAFMOCD w/AFM
Top120119.2121
Middle112117.3115
Bottom143132.8141
 OCD FittingAFMOCD w/Fitting
σTop1.050.750.35
σMiddle1.580.750.60
σBottom0.780.750.42

The statistical approach described here allows researchers to bring together different measurement techniques that may each have their own best attributes, such as excellent bottom-width response versus sensitivity to sidewall changes or high-throughput versus accuracy. It allows us to combine the uncertainties of two or more single measurements into an uncertainty that is an improvement over individual values. In the future, we will focus on combining results from multiple reference-metrology tools, which may may also yield improvements.

The authors would like to thank Pete Lipscomb and Michael Bishop of International SEMATECH for wafer fabrication. We are grateful to Ron Dixson for AFM reference metrology and to Thomas Germer and Alan Heckert of the National Institute of Standards and Technology (NIST) for RCWA model development and simulation support.


Richard Silver, Bryan Barnes, Nien Fan Zhang
NIST
Gaithersburg, MD

Richard Silver is the project leader for overlay and high-resolution optical metrology. He has presented a number of papers in this and related fields. He was recently chairman for the SPIE Microlithography Metrology Inspection and Process Control Conference and is also the co-chair of the Semiconductor Equipment and Materials International (SEMI) microlithography committee.

Hui Zhou
KT Consulting
Gaithersburg, MD