9 August 2023 Simulation of light propagation through the maternal abdomen: towards transabdominal fetal pulse oximetry
Author Affiliations +
Abstract
As an application of near-infrared spectroscopy, pulse oximetry is widely used to non-invasively measure the arterial oxygen saturation (SpO2). To improve the maternal and fetal outcome during delivery, transabdominal fetal pulse oximetry can be used to measure the fetal SpO2. The layered tissue structure above the fetus, however, is complex and thick. In order to understand the feasibility of transabdominal pulse oximetry, we simulated light propagation through the maternal abdomen. For realistic geometry, we segmented a magnetic resonance imaging (MRI) scan of a pregnant women to create a 3D anatomical model, from which a 3D tetrahedron mesh was generated. Using this mesh, we then simulated photon propagation for 5 wavelengths and a grid of 70 sources and detectors (35 each) on the maternal abdomen above the fetal head with NIRFAST. Finally, depth sensitives were examined with Jacobian (J) and flat field. For a fetal head at ~4 cm depth, we found the normalized J at this depth is ~0.1-1% for source-detector distances within 9-10 cm. We also observed that at the same depth, the normalized flat field sensitivity is ~5-10%, which is 1-2 orders of magnitude higher than the normalized J. These results indicate that enough light can reach the fetus when considering ~9 cm source detector distances.
© (2023) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jingyi Wu, Alexander Ruesch, Gopika Satish, Erin Anand, Martin Debreczeny, Eric Johansson, Neil Ray, and Jana M. Kainerstorfer "Simulation of light propagation through the maternal abdomen: towards transabdominal fetal pulse oximetry", Proc. SPIE 12628, Diffuse Optical Spectroscopy and Imaging IX, 126281E (9 August 2023); https://doi.org/10.1117/12.2671002
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Fetus

3D modeling

Abdomen

Oximetry

Magnetic resonance imaging

Brain

Tissues

RELATED CONTENT

Diffuse optical-MRI fusion and applications
Proceedings of SPIE (February 12 2008)
Simulation of MRI with field nonuniformities
Proceedings of SPIE (July 24 1998)
An MRI-guided PET partial volume correction method
Proceedings of SPIE (March 27 2009)

Back to Top