In vivo optical analysis of pancreatic cancer tissue in living model mice using fluorescence and Raman spectroscopic techniques
Author(s):
Toshiaki Suzuki;
Yusuke Hattori;
Takashi Katagiri;
Hiroki Mitsuoka;
Ken-ichi Sato;
Toru Asakura;
Toru Shimosegawa;
Hidetoshi Sato
Show Abstract
Living pancreatic cancer tissues grown subcutaneously in nude mice are studied by in vivo Raman spectroscopy and
autofluorescence imaging. Comparing the same point spectra of alive pancreatic cancer tissue to that of the dead tissue, it
is found that they are different each other. The results suggest that the spectral changes reflect the protein conformational
changes in the tumor tissue with death of the host animal. From the result of autofluorescence study, in vivo
autofluorescence imaging has potential as a method to assign the histological elements of the pancreatic cancer tissue
without any staining. These results strongly suggest that combination of these techniques is very important to study
biological tissue.
Label-free imaging of arterial tissues using photonic crystal fiber (PCF) based nonlinear optical microscopic system
Author(s):
Alex C. T. Ko;
Andrew Ridsdale;
Adrian F. Pegoraro;
Michael S. D. Smith;
Leila B. Mostaço-Guidolin;
Mark D. Hewko;
Elicia M. Kohlenberg;
Bernie J. Schattka;
Masashi Shiomi;
Albert Stolow;
Michael G. Sowa
Show Abstract
Nonlinear optical (NLO) microscopy provides a minimally invasive optical method for
fast molecular imaging at subcellular resolution with 3D sectioning capability in thick,
highly scattering biological tissues. In the current study, we demonstrate the imaging
of arterial tissue using a nonlinear optical microscope based on photonic crystal fiber
and a single femto-second oscillator operating at 800nm. This NLO microscope system
is capable of simultaneous imaging extracellular elastin/collagen structures and lipid
distribution within aortic tissue obtained from coronary atherosclerosis-prone WHHLMI
rabbits (Watanabe heritable hyperlipidemic rabbit-myocardial infarction) Clear
pathological differences in arterial lumen surface were observed between healthy
arterial tissue and atherosclerotic lesions through NLO imaging.
Spectral histology of breast tissue using mid-infrared spectroscopic imaging
Author(s):
F. Nell Pounder;
Rohit Bhargava
Show Abstract
Histopathologic recognition is the gold standard in breast cancer diagnoses and is a primary determinant tool for cancer
research. Unfortunately, the manual nature of histopathologic recognition leads to low throughput analysis, delays in
decision-making and errors. Here, we present an automated means to accurate histologic recognition using mid-infrared
molecular spectroscopy. Fourier transform infrared (FT-IR) spectroscopic imaging is combined with statistical pattern
recognition and high throughput sampling to provide automated tissue segmentation into constituent cell types. The
method does not need dyes or probes and dispenses with human input. Results demonstrate that the technique is capable
of accurate histologic segmentation that can potentially become competitive with that attained by conventional
immunohistochemical analyses.
Full pupil line-scanning confocal microscope for imaging weakly scattering tissues: comparison to divided pupil
Author(s):
Daniel S. Gareau;
Sanjee Abeytunge;
Milind Rajadhyaksha
Show Abstract
Confocal reflectance full-pupil and divided-pupil line-scanning microscopes
provide optical sectioning and image nuclear detail in skin.
Line-scanning with linear
detectors is a simpler alternative to point-scanning for imaging weakly scattering
epidermis and the oral epithelium. With illumination of 830 nm, a water immersion lens
of numerical aperture 0.9 and slit width three times smaller than the diffraction-limited
line width, the instrumental full width at half maximum (FWHM) optical sectioning (linespread
function) for the full-pupil design is 1.4 +/- 0.07 μm, which degrades through fullthickness
human epidermis to 2.8 +/- 0.78 μm. The lateral resolution is 0.7±0.10 μm,
which degrades to 1.6±0.28 μm through human epidermis. The
divided-pupil design
demonstrates instrumental optical sectioning of 1.7 μm, which degrades to 7.6 μm
through human epidermis. The lateral resolution is 1.0 μm, which degrades to 1.7 μm.
Heavy scattering in the dermis decreases contrast. Images of skin
in-vivo show nuclear
detail as expected with the predicted and experimentally verified sectioning. However,
pixel crosstalk and speckle artifact degrade image quality in strongly scattering and
aberrating tissues. The sources of degradation (aberration and scattering) are
evaluated for the two design to assess the feasibility of these techniques for in vivo
imaging.
Fluorescence intensity, lifetime, and anisotropy screening of living cells based on total internal reflection techniques
Author(s):
Thomas Bruns;
Brigitte Angres;
Heiko Steuer;
Wolfgang S. L. Strauss;
Herbert Schneckenburger
Show Abstract
A setup for fluorescence measurements of surfaces of biological samples, in particular the plasma membrane of living
cells, is described. The method is based on splitting of a laser beam and multiple total internal reflections (TIR) within
the bottom of a microtiter plate, such that up to 96 individual samples are illuminated simultaneously by an evanescent
electromagnetic field. Two different screening procedures for the detection of fluorescence arising from the plasma
membrane of living cells by High Throughput Screening (HTS) and High Content Screening (HCS), are distinguished. In
the first case a rapid measurement of large sample numbers based on fluorescence intensity, and in the second case a
high content of information from a single sample based on the parameters fluorescence lifetime (Fluorescence Lifetime
Screening, FLiS) and fluorescence anisotropy (Fluorescence Lifetime Polarization Screening, FLiPS) is achieved. Both
screening systems were validated using cultivated cells incubated with different fluorescent markers (e. g. NBD-cholesterol)
as well as stably transfected cells expressing a fluorescent membrane-associating protein. In addition,
particularly with regard of potential pharmaceutical applications, the kinetics of the intracellular translocation of a
fluorescent protein kinase c fusion protein upon stimulation of the cells was determined. Further, a caspase sensor based
on Förster Resonance Energy Transfer (FRET) between fluorescent proteins was tested. Enhanced cyan fluorescent
protein (ECFP) anchored to the inner leaflet of the plasma membrane of living cells transfers its excitation energy via a
spacer (DEVD) to an enhanced yellow fluorescent protein (EYFP). Upon apoptosis DEVD is cleaved, and energy
transfer is disrupted, as proven by changes in fluorescence intensity and decay times.
Imaging living hair cells within the cochlear epithelium of mice using two-photon microscopy
Author(s):
Tao Yuan;
Simon S. Gao;
Peter Saggau;
John S. Oghalai
Show Abstract
Mice are an excellent model for studying mammalian hearing and transgenic mouse models of human hearing loss
are commonly available for research. However, the mouse cochlea is substantially smaller than other animal models
routinely used to study cochlear physiology. This makes the study of their hair cells difficult. We developed a novel
methodology to optically image calcium within living hair cells left undisturbed within the excised mouse cochlea.
Fresh cochleae were harvested, left intact within their otic capsule bone, and glued upright in a recording chamber. The
bone overlying the region of the cochlear epithelium to be studied was opened and Reissner's membrane was incised. A
fluorescent indicator was applied to the preparation to image intracellular calcium. A custom-built upright two-photon
microscope was used to image the preparation using three dimensional scanning. We were able to image about 1/3 of a
cochlear turn simultaneously, in either the apical or basal regions. Within one hour of animal sacrifice, we found that
outer hair cells demonstrated increased fluorescence compared with surrounding supporting cells. Thus, this
methodology can be used to visualize hair cell calcium changes and mechanotransduction over a region of the epithelium.
Because the epithelium is left within the cochlea, dissection trauma is minimized and artifactual changes in hair cell
physiology are reduced.
Cellular imagery with total internal reflection holographic microscopy
Author(s):
William M. Ash III;
Myung K. Kim
Show Abstract
Evanescent wave phase profiling is incorporated with digital holography into a new surface imaging technique termed
total internal reflection holographic microscopy (TIRHM). Quantitative images of tissue structures and profiles of
cellular membranes are presented to demonstrate the method's performance capabilities. Applications of this technique
include measurements of cellular membranes and their transport processes without the addition of fluorophores. The
angular spectrum method to compensate for tilt anamorphism due to the inclined TIR plane is also discussed.
Spectral imaging of microvascular function in a renal cell carcinoma after treatment with a vascular disrupting agent
Author(s):
Mamta Wankhede;
Casey deDeugd;
Dietmar W. Siemann;
Brian S. Sorg
Show Abstract
Tumors are highly metabolically active and thus require ample oxygen and nutrients to proliferate. Neovasculature
generated by angiogenesis is required for tumors to grow beyond a size of about 1-2mm. Functional tumor vasculature
also provides an access point for development of distant metastases. Due to the importance of the microvasculature for
tumor growth, proliferation, and metastasis, the microvasculature has emerged as a therapeutic target for treatment of
solid tumors. We employed spectral imaging in a rodent window chamber model to observe and measure the oxygen
transport function of tumor microvasculature in a human renal cell carcinoma after treatment with a fast acting vascular
disrupting agent. Human Caki-1 cells were grown in a dorsal skin-fold window chamber in athymic nude mice.
Spectral imaging was used to measure hemoglobin saturation immediately before, immediately after and also at 2, 4, 6,
8, 24 and 48 hours after administration of the tubulin binding agent OXi4503. Up to 4 hours after treatment, tumor
microvasculature was disrupted from the tumor core towards the periphery as seen in deoxygenation as well as
structural changes of the vasculature. Reoxygenation and neovascularization commenced from the periphery towards
the core from 6 - 48 hours after treatment. The timing of the effects of vascular disrupting agents can influence
scheduling of repeat treatments and combinatorial treatments such as chemotherapy and radiation therapy. Spectral
imaging can potentially provide this information in certain laboratory models from endogenous signals with microvessel
resolution.
Structural characterization of colored human iridal melanosomes by photo emission electron microscopy
Author(s):
Dana N. Peles;
Lian Hong;
John D. Simon;
Dan-Ning Hu M.D.
Show Abstract
Ocular uveal melanosomes contain both eumelanin and pheomelanin. The ratio of these two melanins has been discussed
in relation to the epidemiological data for skin cancer rates, with increased incidence observed for increased relative
concentrations of pheomelanin. Recent studies suggest that a similar trend exists underlying the epidemiology of uveal
melanomas. In the present study, the biomolecular organization of human iridal melanosomes from different colored
irises were examined to determine if the photoreactivity changes with the altered eumelanin:pheomelanin ratio, and
whether such changes can account for epidemiological results. Specifically, photoemission electron microscopy
(PEEM), a unique surface-sensitive, direct-imaging technique capable of providing chemical information not obtained
by other electron microscopies, was used in combination with Duke University's tunable UV free electron laser (FEL) to
determine the surface electrochemical properties of melanosomes from blue and dark brown irides. The results
demonstrate that the melanins are organized such that pheomelanin is encased by eumelanin. This "casing model" is
consistent with kinetic information available on the early steps of melanogenesis and provides new insights into
molecular mechanisms underlying the epidemiology of uveal melanoma.
Direct Raman imaging spectroscopy of lung cancer cells and apoptotic cells
Author(s):
Yusuke Oshima;
Chie Furihata;
Hidetoshi Sato
Show Abstract
A Raman spectroscopic technique enables to observe intracellular molecules without fixation or labeling procedures in
situ. We demonstrated a classification of human lung cancer cells with Raman spectroscopy and principal component
analysis. Normal lung cell-lines and 4 pathological types of cancer cell-lines were seeded on culture dishes and
examined. It was as a preliminary study for direct Raman imaging spectroscopy, which could be available for clinical use,
to diagnose cancer. The result suggests that Raman spectroscopy could be a complementary method for
immunohistology study. We also constructed a new direct Raman imaging system consisting of a high sensitive CCD
image sensor, narrow band pass-filters, and a background-free electrically tunable Ti:Sapphire laser. The observation
wavelengths can be switched immediately for the purpose of malignancy rapid diagnosis or real time measurement for
apoptotic cells. The potential ability of the direct Raman imaging system is supposed to evaluate apoptosis by UV
irradiation and anticancer drug-treatment for living lung cancer cells.
Functional imaging of a single cell: far-field infrared super-resolution microscopy using autofluorescence detection
Author(s):
Tsutomu Ohmori;
Keiichi Inoue;
Makoto Sakai;
Masaaki Fujii;
Miya Ishihara M.D.;
Makoto Kikuchi M.D.
Show Abstract
We demonstrated cell imaging without any stain by far-field 2-color infrared (IR) super-resolution microscopy,
combining laser fluorescence microscope and picosecond transient fluorescence detected IR (TFD-IR) spectroscopy.
TFD-IR spectroscopy detects IR absorption by monitoring fluorescence due to an electronic transition from a vibrational
excited level by an additional visible light. By using the IR microscopy based on TFD-IR spectroscopy, the spatial
resolution of the image can be increased to the visible diffraction limit of sub-μm, i.e., the IR is super-resolved. Cell
auto-fluorescence due to flavin molecules was monitored for label-free detection of the cellular components. The
fluorescence image of an A549 cell was obtained by introducing both an IR light at 3300 nm and a visible light at 560
nm. The spatial resolution of the image was estimated to be 1.6 μm. This is about 2.5-times higher resolution than the
diffraction limit of IR light. The fluorescence intensity of the images at 3448 nm was smaller than that at 3300 nm,
corresponding to the smaller IR absorption. Therefore, IR spectral imaging of a single cell was achieved with superresolution.
Physical mechanism of delayed luminescence from human serum
Author(s):
Hua Bai;
Ping Chen;
Lie Lin;
Shengjiang Chang;
Guoqing Tang;
Guoguang Mu
Show Abstract
The fluorescence spectra, delayed luminescence (DL) spectra and DL decay dynamics of human serum were studied
by fluorescence and time resolved emission spectrum technology under different excitation conditions in this paper. The
results we obtained are shown as follows: (1) the DL spectrum is similar to the time resolved fluorescence spectrum
within 50ns after Ps laser pulse excitation. (2) The intensity and decay time of DL from the serum samples are dependent
on excitation power and irradiation time. Under fixed excitation power, the longer irradiation time is, the higher the DL
intensity; after the excitation energy reaches about 200mJ, the DL intensity is nearly unchanged. The change of DL decay
time follows the similar regulation to that of DL intensity. (3) As the excitation energy increases, the spectral distribution
of the relative intensities exhibits an observable change. The higher the excitation energy is, the stronger the relative
intensity at short wavelength region. The results show that the delayed luminescence of human serum is mainly
originated from its delayed fluorescence, phosphorescence, and induced bio-photon emission. These results may be also
useful for the development of serum diagnosis technology.
Evaluation of optimal DNA staining for triggering by scanning fluorescence microscopy (SFM)
Author(s):
Anja Mittag;
Monika Marecka;
Arkadiusz Pierzchalski;
Wolf Malkusch;
József Bocsi;
Attila Tárnok
Show Abstract
In imaging and flow cytometry, DNA staining is a common trigger signal for cell identification. Selection of
the proper DNA dye is restricted by the hardware configuration of the instrument. The Zeiss Imaging Solutions
GmbH (München, Germany) introduced a new automated scanning fluorescence microscope - SFM (Axio
Imager.Z1) which combines fluorescence imaging with cytometric parameters measurement. The aim of the
study was to select optimal DNA dyes as trigger signal in leukocyte detection and subsequent cytometric
analysis of double-labeled leukocytes by SFM.
Seven DNA dyes (DAPI, Hoechst 33258, Hoechst 33342, POPO-3, PI,
7-AAD, and TOPRO-3) were tested and
found to be suitable for the implemented filtersets (fs) of the SFM (fs: 49, fs: 44, fs: 20). EDTA blood was
stained after erythrocyte lysis with DNA dye. Cells were transferred on microscopic slides and embedded in
fluorescent mounting medium. Quality of DNA fluorescence signal as well as spillover signals were analyzed
by SFM. CD45-APC and CD3-PE as well as CD4-FITC and CD8-APC were selected for immunophenotyping
and used in combination with Hoechst.
Within the tested DNA dyes DAPI showed relatively low spillover and the best CV value. Due to the low
spillover of UV DNA dyes a triple staining of Hoechst and APC and PE (or APC and FITC, respectively) could
be analyzed without difficulty. These results were confirmed by FCM measurements.
DNA fluorescence is applicable for identifying and triggering leukocytes in SFM analyses. Although some
DNA dyes exhibit strong spillover in other fluorescence channels, it was possible to immunophenotype
leukocytes. DAPI seems to be best suitable for use in the SFM system and will be used in protocol setups as
primary parameter.
High-performance dispersive Raman and absorption spectroscopy as tools for drug identification
Author(s):
Olga Pawluczyk;
Sam Andrey;
Paul Nogas;
Andrew Roy;
Romuald Pawluczyk
Show Abstract
Due to increasing availability of pharmaceuticals from many sources, a need is growing to quickly and efficiently
analyze substances in terms of the consistency and accuracy of their chemical composition. Differences in chemical
composition occur at very low concentrations, so that highly sensitive analytical methods become crucial. Recent
progress in dispersive spectroscopy with the use of 2-dimensional detector arrays, permits for signal integration along a
long (up to 12 mm long) entrance slit of a spectrometer, thereby increasing signal to noise ratio and improving the ability
to detect small concentration changes. This is achieved with a non-scanning, non-destructive system. Two different
methods using P&P Optica high performance spectrometers were used. High performance optical dispersion Raman and
high performance optical absorption spectroscopy were employed to differentiate various acetaminophen-containing
drugs, such as Tylenol and other generic brands, which differ in their ingredients. A 785 nm excitation wavelength
was used in Raman measurements and strong Raman signals were observed in the spectral range 300-1800 cm-1.
Measurements with the absorption spectrometer were performed in the wavelength range 620-1020 nm. Both Raman and
absorption techniques used transmission light spectrometers with volume phase holographic gratings and provided
sufficient spectral differences, often structural, allowing for drug differentiation.
Label-free optical detection of small-molecule compound microarrays immobilized on solid support using macromolecular scaffolds and subsequent protein binding reactions
Author(s):
Y. S. Sun;
J. P. Landry;
Y. Y. Fei;
X. D. Zhu;
J. T. Luo;
X. B. Wang;
K. S. Lam
Show Abstract
Small-molecule microarrays composed of tens of thousands of distinct synthetic molecules, natural products, and their
combinations/modifications provide a high-throughput platform for studying protein-ligand interactions. Immobilization
of small molecule compounds on solid supports remains a challenge as widely varied small molecules generally lack
unique chemical groups that readily react with singly or even multiply functionalized solid support. We explored two
strategies for immobilizing small molecule compounds on
epoxy-functionalized glass surface using primary-aminecontaining
macromolecular scaffolds: bovine serum albumin (BSA) and
amine-modified poly-vinyl alcohol (PVA).
Small molecules with N-hydroxysuccinimide (NHS) groups were conjugated to BSA or amine-modified PVA.
Small-molecule-BSA conjugates and small-molecule-PVA conjugates were subsequently immobilized on epoxy-functionalized
glass slides through amine-epoxy reactions. Using an
oblique-incidence reflectivity difference (OI-RD) scanning
microscope as a label-free detector, we performed a comparative study of the effectiveness of BSA and PVA as
macromolecular scaffolds for anchoring small molecule compounds in terms of conjugation efficiency, surface
immobilization efficiency, effect of the scaffold on end-point and kinetics of subsequent binding reactions with protein
probes.
Label-free screening small-molecule compound libraries for protein-ligands using a high-throughput optical scanning microscope
Author(s):
Y. Y. Fei;
J. P. Landry;
Y. S. Sun;
X. D. Zhu;
X. B. Wang;
J. T. Luo;
C. Y. Wu;
K. S. Lam
Show Abstract
We describe a new oblique-incidence reflectivity difference (OI-RD) scanning microscope for high-throughput
screening, in microarray format on functionalized glass slides, small-molecule compound libraries for protein ligands.
The microscope employs a combination of scan mirror for y-scan and single-axis translation stage for x-scan. For a
printed microarray with over 10,000 features, each of 100 μm in diameter and distinct small molecule targets, we can
acquire an end-point image of the microarray in 90 minutes with pixel resolution of 20 μm × 20 μm. The microscope is
also capable of measuring binding kinetics of over 10,000
protein-ligand reactions simultaneously. We also describe a
number of strategies for immobilizing small molecule compounds on functionalized glass slides: (1) conjugating the
compounds (through a chemically inert linker) with a lysine residue so that the primary amine on the lysine serves as the
anchor to epoxy-functionalized glass surface; (2) conjugating the compounds (through a linker) with a biotin residue so
that the biotin serves as the anchor to streptavidin-functionalized glass surface; (3) immobilizing small molecule
compounds without modification on isocyanate-functionalized glass surface through non-specific reaction of
nucleophilic molecular motifs on most bioactive compounds with isocyanate groups. We present preliminary
measurements of protein-small molecule binding reactions using the new microscope and the surface immobilization
strategies.
Hollow fiber: a biophotonic implant for live cells
Author(s):
Oscar F. Silvestre;
Mark D. Holton;
Huw D. Summers;
Paul J. Smith;
Rachel J. Errington
Show Abstract
The technical objective of this study has been to design, build and validate biocompatible hollow fiber implants based on
fluorescence with integrated biophotonics components to enable in fiber kinetic cell based assays. A human
osteosarcoma in vitro cell model fiber system has been established with validation studies to determine in fiber cell
growth, cell cycle analysis and organization in normal and drug treated conditions. The rationale for implant
development have focused on developing benchmark concepts in standard monolayer tissue culture followed by the
development of in vitro hollow fiber designs; encompassing imaging with and without integrated biophotonics.
Furthermore the effect of introducing targetable biosensors into the encapsulated tumor implant such as quantum dots for
informing new detection readouts and possible implant designs have been evaluated. A preliminary micro/macro
imaging approach has been undertaken, that could provide a mean to track distinct morphological changes in cells
growing in a 3D matrix within the fiber which affect the light scattering properties of the implant. Parallel engineering
studies have showed the influence of the optical properties of the fiber polymer wall in all imaging modes. Taken all
together, we show the basic foundation and the opportunities for multi-modal imaging within an in vitro implant format.
Probing the mechanical properties of Drosophila embryo ephitelial cells in vivo by laser nanodissection
Author(s):
Matteo Rauzi;
Eric P. Mottay;
Thomas Lecuit;
Pierre-François Lenne
Show Abstract
We used laser nanodissection to study the nature, magnitude and distribution of forces at cell-cell contacts during tissue
morphogenesis in live Drosophila embryos. We designed two set-ups coupling a near infrared femtosecond laser to
either an inverted fluorescence microscope or a spinning-disk confocal microscope. We show that the developed tools
are able to locally disrupt the acto-myosin network while preserving the integrity of the membranes. With these systems,
we could explore the redistribution of cortical elements and relaxation of cell-cell contacts after local ablation.
Photoablative dilution with pre-enrichment for the clonal isolation of rare cancer cells
Author(s):
Michael D. Zordan;
James F. Leary
Show Abstract
The clonal isolation of rare cells, especially cancer and stem cells, in a population is important to cell biology.
We have demonstrated that the Laser-Enabled Analysis and Processing (LEAP, Cyntellect Inc., San Diego, CA)
instrument can be used to efficiently produce clones by photoablative dilution. The LEAP instrument performs
automated fluorescence imaging and real-time image analysis to classify cells. The instrument also features a pulsed
laser that gives it the ability to purify a sample by eliminating unwanted cells via laser ablation or UV-induced
apoptosis. In photoablative dilution, rare cells are deposited into a multiwell plate at 10 cells per well. Then one
cell is chosen to clone, and the other cells present in the well are eliminated by laser ablation. We have successfully
used LEAP to produce single cell clones in 95% of wells (originally containing 5±2.1 cells/well). While
photoablative dilution is a very effective way of producing clonal cultures, it has a fundamental limitation in the low
number of cells that can be processed. This can be overcome by performing a pre-enrichment to increase the
frequency of the rare cells to be cloned. Another enrichment strategy is flow cytometry based cell sorting. Flow
sorting can provide greater than 104 fold enrichment and cells can be sorted directly into a multiwell plate. With
pre-enrichment, photoablative dilution can be used to clonally isolate rare cells. This is especially important in cases
where the total number of potentially rare cells recovered by first stage enrichment sorting is only 10-200 cells.
Such a situation which would normally preclude second pass sorting for purity by the high-throughput first stage cell
separation technology.
Quantitative phase evaluation of dynamic changes on the cell membrane during laser microsurgery
Author(s):
Lingfeng Yu;
Samarendra Mohanty;
Gangjun Liu;
Suzanne Genc;
Myung K. Kim;
Zhongping Chen;
Michael W. Berns
Show Abstract
Digital holographic microscopy allows determination of dynamic changes in the optical thickness
profile of a transparent object with sub-wavelength accuracy. Here, we report a quantitative phase
laser microsurgery system, which takes advantage of the precise optical manipulation by the laser
microbeam and quantitative phase imaging by digital holographic microscopy with high spatial and
temporal resolution. This system would enable absolute quantitation of localized alteration/damage
to transparent phase objects, such as the cell membrane or intra-cellular structures, being exposed to
the laser microbeam, which was not possible using conventional phase-contrast microscopy.
Cell palpation system for local mechanical properties of a cell with an optically manipulated particle
Author(s):
H. Miyoshi;
T. Sugiura;
K. Minato
Show Abstract
During cell adhesion and migration, a cell forms focal adhesion, which connects cytoskeleton with extracellular matrix
(ECM) through integrin, and applies cytoskeletal force to the ECM through focal adhesion. In the initial phase of cell
adhesion (initial adhesion), protein related to cell adhesion recruits other components to reinforce adhesion force and
grows to focal complex. To study the mechanism of cell adhesion, we focused on relationship between variation of
mechanical property of cell adhesion and related protein for cell adhesion. Especially, we approached by understanding
mechanical property of initial adhesion. To measure this property, we developed a "cell palpation system", which utilizes
optical tweezers to apply mechanical stimulus to a cell and to investigate reactive force. As below, this system gives
information on the mechanical property (membrane support tension) and a time course of the property by using an
optically manipulated microbead through an analysis based on mechanical model of this microbead. To create cell
adhesion between the microbead and cell surface, the microbead was coated with collagen and we investigated the
mechanical property of initial adhesion. And we analyzed the processes in relation to maturation of initial adhesion at a
single molecular level.
Dynamic behaviour of fungi in microfluidics: a comparative study
Author(s):
Marie Held;
Marie Binz;
Clive Edwards;
Dan V. Nicolau
Show Abstract
All multi-cellular living organisms are very complex microfluidics systems, which are assembled 'inside-out', as the
result of a complex 'tug-of-war' process comprising both
feed-forward modules, in particular the program embedded in
the species' DNA, and feed-back processes, in particular the response to external environmental conditions. Living
organisms have to solve a perennial problem: for an environment with limited resources (finite and spatial distribution of
nutrients, geometrical limitations, competition from other species, etc.), finite growth rate and penalties applied to nonoptimal
behaviour, what is the best strategy that will maximise growth and survival of the individual and the species?
Identifying how organisms resolve this problem is very difficult due to the high complexity involved. However,
filamentous fungi and other simple organisms, provide a means to address these issues due to the availability of
microscopic imaging and microfabricated geometries in which the behaviour can be tested. Semiconductor fabrication
(photolithography, RIE etching, and soft lithography techniques) was used here to produce artificial micro-networks,
which are physically and chemically structured 3D microenvironments. Continuous imaging of fungal behaviour in
microfluidics networks required the use of PDMS, which is transparent and O2-permeable.
Using these structures, the dynamic growth behaviour of two fungal species Neurospora crassa and Armillaria mellea
was observed microscopically in real-time. Growth parameters such as the tip extension velocities, branching angles and
branching distances on plain surfaces differed vastly between the species. Features in the micronetworks of a size similar
to the hyphal diameter induced the largest change in growth parameters such as, the largest decrease in branching
distances in N. crassa and a near complete suppression for A. mellea. Despite these fundamental differences, both
species negotiated the structures successfully. These results open the way to asking more fundamental questions, such as
which species use algorithms that are most efficient for solving a particular type of mathematical problem coded in a
network? Can these natural algorithms be used for either better control of fungal growth and metabolism for both
medical and biotechnology applications; or even for new computation paradigms?
Ventricular defibrillation combining DC electrical field and electrical pacing: an optical mapping study
Author(s):
Sai Shruthi Musunuri;
Liang Tang;
Boyoung Joung;
Edward J. Berbari;
Shien-Fong Lin
Show Abstract
Although high voltage direct current (DC) shock is a standard technique to terminate ventricular fibrillation
(VF), it can cause severe pain and tissue damage. The exact effect of the DC electric field, which can depolarize the
heart during VF is still unknown. We hypothesized that low-energy DC field in combination with pacing (pacing+DC)
could terminate VF by affecting the ventricular propagation pattern. In six Langendorff-perfused isolated rabbit hearts
with the ablated sinoatrial (SA) node, the DC field was delivered to the left ventricle (cathode) and right ventricle
(anode). We designed a timed protocol using LabVIEW programming that delivers pacing, DC and pacing+DC stimuli
for two seconds time intervals each. The pacing pulse (with varying pacing cycle length: 300ms-30ms) was delivered to
the apex. Transmembrane voltage was recorded with optical mapping technique for 16 seconds at a sampling rate of
2ms/frame. We crushed the sinoatrial node to reduce the heart rate. The baseline activation appeared to have endocardial
origins with a mean escape ventricular rate of 60 ± 5bpm at baseline. The DC field (30mA-60mA) alone increased the
mean heart rate to 120±5bpm. Although DC alone terminated VF in a few cases, the rate of termination was very low
(6.2%). However, when pacing+DC was applied, it was possible to terminate VF in 34 of 130 episodes in six rabbits.
The rate of successful defibrillation of VF with pacing+DC was significantly higher than that with DC alone (20% vs
6.2%, p<0.01). Pacing alone never terminated the VF. In conclusion, DC field may affect the conduction velocity in
normal condition. Pacing+DC intervention could lead to regularization of VF propagation and eventually to termination.
Further improvement of this approach may offer a higher success rate of defibrillation with lower energy requirements.
Effect of time gating and polarization discrimination of propagating light in turbid media during angular domain imaging (ADI)
Author(s):
Fartash Vasefi;
Eldon Ng;
Bozena Kaminska;
Glenn H. Chapman;
Jeffrey J. L. Carson
Show Abstract
Angular Domain Imaging (ADI) employs an angular filter array to accept photons within a small acceptance angle
along the axis of an aligned laser light source and preferentially reject scattered light. Simulations show that the
accepted photons travel the shortest paths between source and detector and are therefore the earliest to arrive. We
fabricated angular filter arrays using silicon bulk micromachining and found that an array of 60 μm square shape microtunnels
1 cm in length accepted photons within 0.48 degree of axis of the micro-tunnels. This small acceptance angle
rejected most of the scattered light and sub-millimeter resolution targets could be resolved in a few centimeters of turbid
medium with at least six times reduced mean free path. ADI through media with higher scattering coefficients was not
achievable due to unwanted acceptance of late arriving scattered photons. To reject the late arriving photons, we added
time-domain filtration and linear polarization to ADI. The implementation of a time-gated camera, a 780 nm femtosecond
pulsed laser, and linear polarization to our ADI system resulted in improved image contrast. The use of ADI
with time-gating (gate width 250 ps) and linear polarization enabled visualization of sub-millimeter absorbing objects
with approximately eight times higher image contrast compared to ADI in a scattering medium equivalent to six times
reduced mean free path.
Solid state photomultipliers for biomedical imaging applications
Author(s):
Gin-Chung Wang;
Jeffrey Gordon
Show Abstract
Photomultiplier tubes (PMTs) are often used in scanning imaging systems requiring high sensitivity, due to their low
noise and high gain. Solid-state photomultipliers (SSPMs), an array of independent Geiger-mode avalanche photodiodes,
each with an integrated quenching resistor, have shown potential to outperform PMTs in terms of signal to noise ratio
(SNR) because of higher achievable photon detection efficiency and lower excessive noise factor. Here, the factors
affecting SNR of commercially available PMTs and SSPMs will be compared under different wavelengths (simulating
dye emissions: 500-700 nm) in order to quantify the potential performance gain when PMTs are replaced by SSPMs.
High-speed fluorescence lifetime measurement for investigation of dynamic phenomena
Author(s):
Youngjae Won;
Sucbei Moon;
Dongsoo Lee;
Dug Y. Kim
Show Abstract
We present a novel method for high-speed measurements of fluorescence lifetime, in which fluorescence signal for
precise lifetime determination is acquired in a short time on the order of microseconds. Our method is based on analog
signal that contains a number of fluorescence photons in a pulse, on the contrary to the conventional time-correlated
single-photon counting in which only a single photon is permitted for a fluorescence pulse. Because this method does not
have any problem of photon counting pile-up, the measurement speed is not limited by the single-photon constraint and
can increase up to the excitation repetition rate. In order to extract the accurate lifetime information from the analog
signal contaminated by the slow instrumental response function (IRF), we have developed a new signal processing
method, in which the lifetime is determined by difference between mean arrival time of the analog photo-electronic pulse
of fluorescence signal and one of IRF signal. By both experimental and theoretical studies, we have verified that the
measurement accuracy and precision are nearly independent of the width of the IRF so that inexpensive narrowbandwidth
photo-detectors and low-speed electronics can be used for this method. Excellent accuracy and precision have
been obtained experimentally for high-speed measurements completed in a few microseconds. These results suggest that
our method can be well applied to measurement of fast dynamic phenomena and real time fluorescence lifetime imaging
microscope with low cost.
Red blood cell structure and dynamics explored with digital holographic microscopy
Author(s):
P. Marquet;
B Rappaz;
A. Barbul;
R. Korenstein;
C. Depeursinge;
P. Magistretti
Show Abstract
Digital holographic microscopy (DHM) is a technique that allows obtaining, from a single recorded hologram,
quantitative phase image of living cell with interferometric accuracy. Specifically the optical phase shift
induced by the specimen on the transmitted wave front can be regarded as a powerful endogenous contrast
agent, depending on both the thickness and the refractive index of the sample. Thanks to a decoupling
procedure cell thickness and intracellular refractive index can be measured separately. Consequently, Mean
corpuscular volume (MCV) and mean corpuscular hemoglobin concentration (MCHC), two highly relevant
clinical parameters, have been measured non-invasively at a single cell level. The DHM nanometric axial and
microsecond temporal sensitivities have permitted to measure the red blood cell membrane fluctuations (CMF)
on the whole cell surface.
Image contrast enhancement during time-angular domain imaging through turbid media by estimation of background scattered light
Author(s):
Eldon Ng;
Fartash Vasefi;
Bozena Kaminska;
Glenn H. Chapman;
Jeffrey J. L. Carson
Show Abstract
Time-angular domain imaging (TADI) employs an angular filter array, which functions to accept quasi-ballistic photons
with trajectories near the axis of a collimated light source. At high scattering coefficients, image contrast declines due to
background signals from scattered photons that have trajectories compatible with the angular filter array. We attempted
to correct for the background signal using a temporal discrimination technique and image subtraction. During TADI
through turbid media, photons at early arrival times represent a mixture of quasi-ballistic and scattered photons, while
late arriving photons represent scattered photons. We captured two TADI images of a resolution target suspended
midway through a 2 cm thick cuvette filled with 0.30% IntralipidTM. A 780 nm, 100 ps pulsed laser (PicoTA,
PicoQuant) was used to trans-illuminate the cuvette. Detection was performed after the angular filter array (500
elements with 60 μm × 60 μm square-shaped cross section and 1 cm length) with a gated camera (Picostar HR,
LaVision). The first TADI image was collected at a short gate delay with respect to the minimum transit time, and
resulted in a projection of the target. A long gate delay was used to collect the second TADI image and the projection of
the target was not apparent. A corrected image (two - one) was digitally computed. Analysis of the first image compared
to the corrected image revealed a 2.1-fold increase in
contrast-to-noise ratio for the corrected image. Therefore, images
collected with TADI were improved by processing successive images at different gate delays.
Dynamic reconfigurable spectral imaging microscopy
Author(s):
Bing Chen;
Jame J. Yang;
Michael R. Wang
Show Abstract
We report our recent introduction of a dynamic spectral imaging technique into a microscope system for specimen
examination and classification. It uses a dynamic spectral filter and a diffraction grating on an intermediate image plane
for fast spectral image acquisition that only captures interested spectral image frames. The reconfigurable capability
enables the optimal spectral band selection for a specific application without hardware component modification. It can
achieve 8 nm fine spectral resolution in the whole visible spectral band suitable for fluorescent image evaluation. With a
fast spectral imaging update, it can support large quantify specimen screening and classification.
Hyperspectral microscopy imaging to analyze pathology samples with multicolors reduces time and cost
Author(s):
Michael L. Huebschman;
Kevin P. Rosenblatt;
Harold R. Garner
Show Abstract
By taking advantage of the continuous spectrum collected for each image pixel by the Hyperspectral Microscopy
Imaging (HMI) system, the data spectra can be de-convolved into a set of standard spectra coefficients for each pixel.
The coefficients are calculated by the analysis program and indicate the relative amounts of each standard necessary to
reproduce the data spectra for each image pixel. Images of a single color or composite images of two or more colors can
be produced for visual analysis. The HMI system is composed of an Olympus inverted microscope, imaging
spectrograph, CCD camera, motorized X-Y stage and illumination sources. This system has been used to scan and
analyze pathology tissue samples which have been stained with 4 standard fluorochromes (not including the dynamic
background spectrum) attached to specific antibodies. The typical wavelength range is 420nm to 785nm with the
longest wavelength markers emitting in the spectral region where the human eye is not sensitive. Quantitative values are
recorded and can be compared to the visual interpretations.
Microscopy with spatial filtering for sorting particles and monitoring subcellular morphology
Author(s):
Jing-Yi Zheng;
Zhen Qian;
Robert M. Pasternack;
Nada N. Boustany
Show Abstract
Optical scatter imaging (OSI) was developed to non-invasively track real-time changes in particle morphology with submicron
sensitivity in situ without exogenous labeling, cell fixing, or organelle isolation. For spherical particles, the
intensity ratio of wide-to-narrow angle scatter (OSIR, Optical Scatter Image Ratio) was shown to decrease monotonically
with diameter and agree with Mie theory. In living cells, we recently reported this technique is able to detect
mitochondrial morphological alterations, which were mediated by the Bcl-xL transmembrane domain, and could not be
observed by fluorescence or differential interference contrast images. Here we further extend the ability of morphology
assessment by adopting a digital micromirror device (DMD) for Fourier filtering. When placed in the Fourier plane the
DMD can be used to select scattering intensities at desired combination of scattering angles. We designed an optical
filter bank consisting of Gabor-like filters with various scales and rotations based on Gabor filters, which have been
widely used for localization of spatial and frequency information in digital images and texture analysis. Using a model
system consisting of mixtures of polystyrene spheres and bacteria, we show how this system can be used to sort particles
on a microscopic slide based on their size, orientation and aspect ratio. We are currently applying this technique to
characterize the morphology of subcellular organelles to help understand fundamental biological processes.
Molecular analysis of arterial remodeling: a novel application of infrared imaging
Author(s):
Brad C. Herman;
Rishi Kundi;
Dai Yamanouchi;
K. Craig Kent;
Bo Liu;
Nancy Pleshko
Show Abstract
Arterial remodeling, i.e. changes in size and/or structure of arteries, plays an important role in vascular disease.
Conflicting findings have been reported as to whether an abundance of collagen causes inward or outward remodeling,
phenomena that result in either a smaller or larger lumen, respectively. We hypothesize that the amount, type and quality
of collagen influence the remodeling response. Here, we create mechanical injury to the rat carotid artery using a balloon
catheter, and this leads to inward remodeling. Treatment of the artery with Connective Tissue Growth Factor (CTGF)
causes outward remodeling. We investigated the arterial composition in injured CTGF-treated and non-CTGF-treated
and sham CTGF-treated and non-CTGF treated arteries 14 days post-injury (n = 7-8 per group) using infrared imaging.
A Perkin Elmer Spotlight Spectrum 300 FT-IR microscope was used for data collection. Cross-sections of paraffinembedded
arteries were scanned at 2 cm-1 spectral resolution with spatial resolution of 6.25 μm/pixel, and data analyzed
using Malvern Instruments ISys 5.0. Post-injury, we found a nearly 50% reduction in the average 1338/AM2 area ratio
(correlated to collagen helical integrity). The most dramatic change was a 600% increase in the 1660/1690 peak height
ratio, which has previously been related to collagen crosslink maturity. In all cases, CTGF treatment resulted in the
observed changes in peak parameters normalized back to control values. Overall, these preliminary studies demonstrate
that infrared imaging can provide insight into the underlying molecular changes that contribute to arterial disease.
Measurements and analysis in imaging for biomedical applications
Author(s):
Timothy L. Hoeller
Show Abstract
A Total Quality Management (TQM) approach can be used to analyze data from biomedical optical and imaging
platforms of tissues. A shift from individuals to teams, partnerships, and total participation are necessary from health
care groups for improved prognostics using measurement analysis. Proprietary measurement analysis software is
available for calibrated, pixel-to-pixel measurements of angles and distances in digital images. Feature size, count, and
color are determinable on an absolute and comparative basis. Although changes in images of histomics are based on
complex and numerous factors, the variation of changes in imaging analysis to correlations of time, extent, and
progression of illness can be derived. Statistical methods are preferred. Applications of the proprietary measurement
software are available for any imaging platform. Quantification of results provides improved categorization of illness
towards better health.
As health care practitioners try to use quantified measurement data for patient diagnosis, the techniques reported can be
used to track and isolate causes better. Comparisons, norms, and trends are available from processing of measurement
data which is obtained easily and quickly from Scientific Software and methods. Example results for the class actions of
Preventative and Corrective Care in Ophthalmology and Dermatology, respectively, are provided. Improved and
quantified diagnosis can lead to better health and lower costs associated with health care. Systems support improvements
towards Lean and Six Sigma affecting all branches of biology and medicine.
As an example for use of statistics, the major types of variation involving a study of Bone Mineral Density (BMD) are
examined. Typically, special causes in medicine relate to illness and activities; whereas, common causes are known to be
associated with gender, race, size, and genetic make-up. Such a strategy of Continuous Process Improvement (CPI)
involves comparison of patient results to baseline data using F-statistics. Self-parings over time are also useful. Special
and common causes are identified apart from aging in applying the statistical methods. In the future, implementation of
imaging measurement methods by research staff, doctors, and concerned patient partners result in improved health
diagnosis, reporting, and cause determination. The long-term prospects for quantified measurements are better quality in
imaging analysis with applications of higher utility for heath care providers.
Determination of glucose concentration using Fourier domain optical coherence tomogram
Author(s):
Yasser H. El-Sharkawy
Show Abstract
In order to enhance cell culture growth in biosensors such as those for glucose detection must be developed that are capable
of monitoring cell culture processes continuously and accurate. Fourier domain optical coherence tomography (FD-OCT) is
used to obtain cell images with nanometer level resolution by analyzing the interference pattern by the mixing of reference
and objective light to determine glucose concentration in doped double distilled water and create a glucose signature
spectrum in salt-sugar solution. We demonstrate ultrahigh-resolution optical coherence tomography (OCT) imaging of in
vitro biological cells and an improved deflection angle measurements formal and back projection method is used to
reconstruct the two-dimensional glucose concentration performs refractive index distribution. Slopes of OCT signals
decreased substantially and almost linearly with the increase of glucose concentration from 2.5 to 15 mg/dl. Phantom studies
demonstrated 1% accuracy of scattering- coefficient measurement. Our theoretical and experimental studies suggest that
glucose concentration can potentially be measured non-invasively with high sensitivity and accuracy with OCT systems.
A two-stage morphological classifier of foci occurring in cell transformation assays
Author(s):
Giovanni F. Crosta;
Chiara Urani;
Luca Bussinelli
Show Abstract
Cell Transformation Assays (CTA) rely on the detection of phenotypic changes, namely foci, induced by chemicals (e.g.,
xenobiotics or candidate drugs) in mammalian cells such as C3H10T1/2 mouse fibroblasts. A focus is a cell colony and
as such is made visible by standardized techniques of light microscopy. Foci exhibit a variety of morphological features,
by which three "Types" have been defined. Types II and III consist of cells having undergone neoplastic transformation.
The assignment of a focus to a Type is based on the evaluation of phenotypic features by a trained human expert. An
automated, two-stage morphological classifier of foci is described herewith. Morphological descriptors are extracted
from light microscope images by the "spectrum enhancement" algorithm, which separates structure from texture. Said
descriptors are submitted to a classifier, the first stage of which is trained to discriminate transformed cells from normal
ones and the 2nd stage to discriminate Type III from Type II. The classifier operating in recognition mode (on images not
used for training) is satisfactory in terms of confusion matrix entries. The whole procedure is aimed at removing
subjectivity from the scoring and classification of foci and thus make CTA a more powerful tool in carcinogenesis
studies.
Dynamic fluorescence imaging for multiparametric measurement of tumor vasculature
Author(s):
Myunghwan Choi;
Taesu Chung;
Kyungsun Choi;
Chulhee Choi
Show Abstract
Angiogenesis is essential for tumor growth and a promising target for cancer therapy. Blood vessel monitoring is an
indispensable tool for evaluation and development of anti-angiogenic drugs. Here, we report a new noninvasive in vivo
imaging tool, named dynamic fluorescence imaging (DyFI), for the simultaneous measurement of multiple vascular
parameters, such as vascular density, perfusion rate, and permeability, using spatiotemporal profiles of indocyanine green.
Using DyFI in a tumor xenograft model, we quantitatively measured multiple vascular parameters in tumors and normal
tissues with high spatial resolution. The multimodality of this method allowed us to find a negative spatial correlation
between perfusion and permeability. Moreover, DyFI was effective for revealing the early effects of an anti-angiogenic
drug; these findings were validated using two-photon microscopy. We suggest that DyFI could be a useful tool for the
preclinical development of anti-angiogenic drugs.
The measurement of red blood cell volume induced by Ca2+ based on full field quantitative phase microscopy
Author(s):
Seungrag Lee;
Ji Yong Lee;
Wenzhong Yang;
Dug Young Kim
Show Abstract
We present the measurement of red blood cell (RBC) volume change induced by Ca2+ for a live cell imaging with
full field quantitative phase microscopy (FFQPM). FFQPM is based on the Mach-Zehnder interferometer combined with
an inverted microscopy system. We present the effective method to obtain a clear image and an accurate volume of the
cells. An edge detection technique is used to accurately resolve the boundary between the cell line and the suspension
medium. The measurement of the polystyrene bead diameter and volume has been demonstrated the validity of our
proposed method. The measured phase profile can be easily converted into thickness profile. The measured polystyrene
bead volume and the simulated result are about 14.74 μm3 and 14.14 μm3, respectively. The experimental results of our proposed method agree well with the simulated results within less than 4 %. We have also measured the volume
variation of a single RBC on a millisecond time scale. Its mean volume is 54.02 μm3 and its standard deviation is 0.52
μm3. With the proposed system, the shape and volume changes of RBC induced by the increased intracellular Ca2+ are measured after adding ionophore A23187. A discocyte RBC is deformed to a spherocyte due to the increased intracellular Ca2+ in RBC. The volume of the spherocyte is 47.88 μm3 and its standard deviation is 0.19 μm3. We have demonstrated that the volume measurement technique is easy, accurate, and robust method with high volume sensitivity (<0.0000452 μm3) and this provides the ability to study a biological phenomenon in Hematology.
A container for the advanced cytometry standard (ACS)
Author(s):
Robert C. Leif;
Josef Spidlen;
Ryan R Brinkman
Show Abstract
Introduction: The highest priority for the Advanced Cytometry Standard (ACS) is the interpretation of list-mode
cytometry measurements. Other priorities of lesser importance are the capacity to reproduce a cytometry measurement
and the implementation of a digital microscopy image standard. The sequential nature of these requirements is being
accommodated by a flexible, modular design. A major feature of this modular design is the creation of a design for an
Advanced Cytometry Standard Container (ACSC) that includes a Table of Contents (ToC) XML file, one or more binary
data containing files and files that contain the meta-data that describes the binary data.
Methods: The design and partial implementation of the CytometryML schemas have been based on the techniques of
modularity (each schema describing one object), iterative (spiral) development, inheritance, and reuse. Data-types
including their definitions have been reused from DICOM, FCS, and other standards.
Results: A prototype ToC schema together with prototypes of many of the schemas that describe the contents of the
ACSC have been created together with their supporting schemas. These schemas have been validated with two tools and
XML pages were generated from the main element(s) of the highest level schemas. These elements describe the table of
contents of the zipped container file and a flow-cytometry instrument. The zipped container file (ACSC) describes and
contains the meta and binary data.
Evaluation of slide based cytometry (SBC) for concentration measurements of fluorescent dyes in solution
Author(s):
Arkadiusz Pierzchalski;
Monika Marecka;
Hans-Willy Müller;
József Bocsi;
Attila Tárnok
Show Abstract
Flow cytometers (FCM) are built for particle measurements. In principle, concentration
measurement of a homogeneous solution is not possible with FCM due to the lack of a trigger
signal. In contrast to FCM slide based cytometry systems could act as tools for the measurement
of concentrations using volume defined cell counting chambers. These chambers enable to
analyze a well defined volume. Sensovation AG (Stockach, Germany) introduced an automated
imaging system that combines imaging with cytometric features analysis. Aim of this study was
to apply this imaging system to quantify the fluorescent molecule concentrations.
The Lumisens (Sensovation AG) slide-based technology based on fluorescence digital imaging
microscopy was used. The instrument is equipped with an inverted microscope, blue and red
LEDs, double band-pass filters and a high-resolution cooled 16-bit digital camera. The
instrument was focussed on the bottom of 400μm deep 6 chamber slides (IBIDI GmbH,
Martinsried, Germany) or flat bottom 96 well plates (Greiner Bio One GmbH, Frickenhausen,
Germany). Fluorescent solutions were imaged under 90% pixel saturation in a broad
concentration range (FITC: 0.0002-250 μg/ml, methylene blue (MethB): 0.0002-250 μg/ml).
Exposition times were recorded. Images were analysed by the iCys (CompuCyte Corp.,
Cambridge, MA, USA) image analysis software with the phantom contour function. Relative
fluorescence intensities were calculated from mean fluorescence intensities per phantom contours
divided by the exposition time.
Solution concentrations could be distinguished over a broad dynamic range of 3.5 to 5.5 decades
log (range FITC: 0.0002-31.25μg/ml, MethB: 0.0076-31.25μg/ml) with a good linear relationship
between dye concentration and relative fluorescence intensity. The minimal number of
fluorescent molecules per pixel as determined by the mean fluorescence intensity and the
molecular weight of the fluorochrome were about 800 molecules FITC and ~2.000 MethB.
The novel slide-based imaging system is suitable for detection of fluorescence differences over a
broad range of concentrations. This approach may lead to novel assays for measuring
concentration differences in cell free solutions and cell cultures e.g. in secretion assays.
Triggering of leukocytes by phase contrast in imaging cytometry with scanning fluorescence microscope (SFM)
Author(s):
József Bocsi;
Arkadiusz Pierzchalski;
Monika Marecka;
Wolf Malkusch;
Attila Tárnok
Show Abstract
Slide-based cytometry (SBC) leads to breakthrough in cytometry of cells in tissues, culture and
suspension. Carl Zeiss Imaging Solutions' new automated SFM combines imaging with cytometry.
A critical step in image analysis is selection of appropriate triggering signal to detect all objects.
Without correct target cell definition analysis is hampered. DNA-staining is among the most
common triggering signals. However, the majority of DNA-dyes yield massive spillover into other
fluorescence channels limiting their application. By microscopy objects of >5μm diameter can be
easily detected by phase-contrast signal (PCS) without any staining. Aim was to establish PCS -
triggering for cell identification.
Axio Imager.Z1 motorized SFM was used (high-resolution digital camera, AxioCam MRm;
AxioVision software: automatic multi-channel scanning, analysis). Leukocytes were stained with
FITC (CD4, CD8) and APC (CD3) labelled antibodies in combinations using whole blood method.
Samples were scanned in three channels (PCS/FITC/APC). Exposition-times for PCS were set as
low as possible; the detection efficiency was verified by fluorescence. CD45-stained leukocytes
were counted and compared to the number of PCS detected events. Leukocyte subtyping was
compared with other cytometers.
In focus the PCS of cells showed ring-form that was not optimal for cell definition. Out of focus
PCS allows more effective qualitative and quantitative cell analyses. PCS was an accurate triggering
signal for leukocytes enabling cell counting and discrimination of leukocytes from platelets.
Leukocyte subpopulation frequencies were comparable to those obtained by other cytometers. In
conclusion PCS is a suitable trigger-signal not interfering with fluorescence detection.
Laser rastering flow cytometry: fast cell counting and identification
Author(s):
G. Vacca;
M. R. Junnarkar;
N. R. Goldblatt;
M. W. Yee;
B. M. Van Slyke;
T. C. Briese
Show Abstract
We describe the concept of laser rastering flow cytometry, where a rapidly scanning laser beam allows counting and
classification of cells at much higher rates than currently possible. Modifications to existing flow cytometers to
implement the concept include an acousto-optic deflector, fast analog-to-digital conversion, and a two-step
digital-signal-processing scheme that handles the high data rates and provides key assay information. Results are shown that
prove the concept, demonstrating the ability to resolve closely spaced cells and to measure cells at rates more than an
order of magnitude faster than on conventional flow-cytometer-based hematology analyzers.
Angiogram, fundus, and oxygen saturation optic nerve head image fusion
Author(s):
Hua Cao;
Bahram Khoobehi
Show Abstract
A novel multi-modality optic nerve head image fusion approach has been successfully designed. The new approach has
been applied on three ophthalmologic modalities: angiogram, fundus, and oxygen saturation retinal optic nerve head
images. It has achieved an excellent result by giving the visualization of fundus or oxygen saturation images with a
complete angiogram overlay. During this study, two contributions have been made in terms of novelty, efficiency, and
accuracy. The first contribution is the automated control point detection algorithm for multi-sensor images. The new
method employs retina vasculature and bifurcation features by identifying the initial good-guess of control points using
the Adaptive Exploratory Algorithm. The second contribution is the heuristic optimization fusion algorithm. In order to
maximize the objective function (Mutual-Pixel-Count), the iteration algorithm adjusts the initial guess of the control
points at the sub-pixel level. A refinement of the parameter set is obtained at the end of each loop, and finally an optimal
fused image is generated at the end of the iteration. It is the first time that Mutual-Pixel-Count concept has been
introduced into biomedical image fusion area. By locking the images in one place, the fused image allows
ophthalmologists to match the same eye over time and get a sense of disease progress and pinpoint surgical tools. The
new algorithm can be easily expanded to human or animals' 3D eye, brain, or body image registration and fusion.
High-resolution imaging of biological cell with fiber-based composite interferometer
Author(s):
Hsiu-I Hsu;
Chun-Wei Chang;
Max T.-K. Hou;
I-Jen Hsu
Show Abstract
We proposed and demonstrated an optical system for high-speed and high-resolution imaging of surface profiles. The
system is a fiber-based composite interferometer in which a Michelson interferometer is used for the measurement of the
surface profile of the sample and a Mach-Zehnder interferometer is used for compensation of phase deviation due to the
systematic errors and environmental perturbations. In the system, a laser diode with wavelength of 531 nm was used as
the light source. A two-axis translation stage was used for lateral scanning of the sample. In the reference arm, a
reflection mirror was arranged on a translation stage driven by a piezoelectric transducer as the axial scanning
component. The phase difference between the interferograms of both interferometers was acquired to obtain the surface
height of the sample. The axial accuracy of ±5 nm was achieved where the imaging was acquired within one minute for a
frame. The lateral resolution was at the diffraction limit of light. The system possesses the advantages of low cost,
portability and flexibility. Furthermore, it can perform
high-resolution imaging in large area without special shielding of
the system as well as any special preparation of the sample.
Atmospheric pressure femtosecond laser imaging mass spectrometry
Author(s):
Yves Coello;
Tissa C. Gunaratne;
Marcos Dantus
Show Abstract
We present a novel imaging mass spectrometry technique that uses femtosecond laser pulses to directly ionize the
sample. The method offers significant advantages over current techniques by eliminating the need of a laser-absorbing
sample matrix, being suitable for atmospheric pressure sampling, and by providing 10μm resolution, as demonstrated
here with a chemical image of vegetable cell walls.
Optical property of red blood cell with a phase microscopy interferometer
Author(s):
Ji Yong Lee;
Seungrag Lee;
Wenzhong Yang;
Dug Young Kim
Show Abstract
We present a novel method to determine the effective elastic constant (EEC) and effective restoring
force (ERF) by using volumetric analysis of Red blood cell (RBC)s with Full field quantitative phase
microscopy (FFQPM). We use the simple harmonic oscillator model to determine EEC and ERF. We
investigate the EECs and ERFs of different shape of RBCs (discocyte, acanocyte, stomatocyte, and
spherocyte) and we investigate the effective temporal coherence of RBCs by analyzing temporal
volumetric behavior of the RBCs.
Polarization microscope by single polarization modulation method: novel instrument for visualizing cytoskeleton
Author(s):
In Hee Shin;
Sang-Mo Shin;
Namhui Ahn;
Dug Young Kim
Show Abstract
A novel and simplest polarization microscope by single polarization modulation method to visualize cytoskeletal
structure in bio-cells have been developed. The polarization microscope visualizes cytoskeletal structure based on
birefringence characteristics of cells. In the polarization microscope, a series of intensity images according to single
modulation of rotating linear polarizer were obtained to visualize cytoskeletal structure. Phase retardation and principal
axis orientation were calculated by analytic data processing from the acquired intensity images. Phase retardation and
principal axis orientation of breast cancer cells, MCF 7 and MDA MB 231 were measured and visualized as their
birefringence characteristics by the polarization microscope. Phase retardation sensitivity of the polarization microscope
system also was measured in this paper.