Share Email Print
cover

Proceedings Paper

Spectral Efficiency and Resolution of Si(Li)-Detectors for Photon Energies between 0.3 keV and 5 keV
Author(s): F. Riehle; E. Tegeler; B. Wende
Format Member Price Non-Member Price
PDF $17.00 $21.00

Paper Abstract

The spectral efficiencies of energy dispersive Si(Li) photon counters have been measured using the storage ring BESSY as a radiometric standard source of extremely low photon flux of the order of 1 photon/(s eV). The detectors were irradiated with white synchrotron radia-tion when the storage ring was operated with only about 5 electrons stored. For energy calibration and measurement of the energy resolution X-ray emission lines excited by a 55-Fe source were used. Towards lower photon energies the efficiency is drastically decreased by the building-up of an ice-layer on the permanently cooled detector. By this surface contamination also the energy resolution of the detector is affected. The performance of detector can be recovered by a warming-up procedure.

Paper Details

Date Published: 1 January 1986
PDF: 6 pages
Proc. SPIE 0733, Soft X-Ray Optics and Technology, (1 January 1986); doi: 10.1117/12.964950
Show Author Affiliations
F. Riehle, Institut Berlin (Germany)
E. Tegeler, Institut Berlin (Germany)
B. Wende, Institut Berlin (Germany)


Published in SPIE Proceedings Vol. 0733:
Soft X-Ray Optics and Technology
E. Koch; Guenther A. Schmahl, Editor(s)

© SPIE. Terms of Use
Back to Top
PREMIUM CONTENT
Sign in to read the full article
Create a free SPIE account to get access to
premium articles and original research
Forgot your username?
close_icon_gray