Proceedings Paper
Validation of a method to convert an image to appear as if acquired using a different digital detectorFormat | Member Price | Non-Member Price |
---|---|---|
$17.00 | $21.00 |
Paper Abstract
A method to convert digital mammograms acquired on one system to appear as if acquired using another system is
presented. This method could be used to compare the clinical efficacy of different systems. The signal transfer properties
modulation transfer function (MTF) and noise power spectra (NPS) were measured for two detectors - a computed
radiography (CR) system and a digital radiography (DR) system. The contributions to the NPS from electronic, quantum
and structure sources were calculated by fitting a polynomial at each spatial frequency across the NPS at each dose. The
conversion process blurs the original image with the ratio of the MTFs in frequency space. Noise with the correct
magnitude and spatial frequency was added to account for differences in the detector response and dose. The method was
tested on images of a CDMAM test object acquired on the two systems at two dose levels. The highest dose images were
converted to lower dose images for the same detector, then images from the DR system were converted to appear as if
acquired at a similar dose using CR. Contrast detail curves using simulated CDMAM images closely matched those of
real images.
Paper Details
Date Published: 16 March 2011
PDF: 12 pages
Proc. SPIE 7961, Medical Imaging 2011: Physics of Medical Imaging, 79614F (16 March 2011); doi: 10.1117/12.878756
Published in SPIE Proceedings Vol. 7961:
Medical Imaging 2011: Physics of Medical Imaging
Norbert J. Pelc; Ehsan Samei; Robert M. Nishikawa, Editor(s)
PDF: 12 pages
Proc. SPIE 7961, Medical Imaging 2011: Physics of Medical Imaging, 79614F (16 March 2011); doi: 10.1117/12.878756
Show Author Affiliations
Alistair Mackenzie, The Royal Surrey County Hospital NHS Foundation Trust (United Kingdom)
Adam Workman, Forster Green Hospital (United Kingdom)
David R. Dance, The Royal Surrey County Hospital NHS Foundation Trust (United Kingdom)
Univ. of Surrey (United Kingdom)
Adam Workman, Forster Green Hospital (United Kingdom)
David R. Dance, The Royal Surrey County Hospital NHS Foundation Trust (United Kingdom)
Univ. of Surrey (United Kingdom)
Mary Yip, Univ. of Surrey (United Kingdom)
Kevin Wells, Univ. of Surrey (United Kingdom)
Kenneth C. Young, The Royal Surrey County Hospital NHS Foundation Trust (United Kingdom)
Univ. of Surrey (United Kingdom)
Kevin Wells, Univ. of Surrey (United Kingdom)
Kenneth C. Young, The Royal Surrey County Hospital NHS Foundation Trust (United Kingdom)
Univ. of Surrey (United Kingdom)
Published in SPIE Proceedings Vol. 7961:
Medical Imaging 2011: Physics of Medical Imaging
Norbert J. Pelc; Ehsan Samei; Robert M. Nishikawa, Editor(s)
© SPIE. Terms of Use