Proceedings Paper
Predicting contrast detail performance from objective measurements in digital mammographyFormat | Member Price | Non-Member Price |
---|---|---|
$17.00 | $21.00 |
Paper Abstract
European Guidelines for quality control in digital mammography specify minimum and achievable standards of image
quality in terms of threshold contrast, based on readings of images of the CDMAM test object by human observers.
However the methodology is time-consuming and has large inter- and intra-observer error. To overcome these problems
a software program is available to automatically read CDMAM images. An alternative approach would be to predict
threshold contrast from measurements of DQE and MTF using a model of the imaging process. A simple signal-matched
noise-integration model has been used to predict the contrast detail response of five different types of commercial digital
mammography system (Siemens Inspiration, GE Senographe DS, and three types of Konica Minolta computerised
radiography system). Measurements were made of the MTF and DQE of each detector and the noise equivalent apertures
calculated. For each system sets of 16 images of the CDMAM test object were acquired at a range of dose levels and
contrast-detail plots obtained using human observers and automated reading. The theoretically and experimentally
determined threshold contrasts were compared. An encouragingly good level of agreement was found between the
experimental data and theoretical predictions.
Paper Details
Date Published: 13 March 2009
PDF: 11 pages
Proc. SPIE 7258, Medical Imaging 2009: Physics of Medical Imaging, 72581C (13 March 2009); doi: 10.1117/12.811616
Published in SPIE Proceedings Vol. 7258:
Medical Imaging 2009: Physics of Medical Imaging
Ehsan Samei; Jiang Hsieh, Editor(s)
PDF: 11 pages
Proc. SPIE 7258, Medical Imaging 2009: Physics of Medical Imaging, 72581C (13 March 2009); doi: 10.1117/12.811616
Show Author Affiliations
Kenneth C. Young, The Royal Surrey County Hospital (United Kingdom)
Univ. of Surrey (United Kingdom)
Abdulaziz Alsager, Univ. of Surrey (United Kingdom)
David R. Dance, The Royal Surrey County Hospital (United Kingdom)
Univ. of Surrey (United Kingdom)
Abdulaziz Alsager, Univ. of Surrey (United Kingdom)
David R. Dance, The Royal Surrey County Hospital (United Kingdom)
Jennifer M. Oduko, The Royal Surrey County Hospital (United Kingdom)
Ozcan Gundogdu, The Royal Surrey County Hospital (United Kingdom)
Nicholas M. Spyrou, Univ. of Surrey (United Kingdom)
Ozcan Gundogdu, The Royal Surrey County Hospital (United Kingdom)
Nicholas M. Spyrou, Univ. of Surrey (United Kingdom)
Published in SPIE Proceedings Vol. 7258:
Medical Imaging 2009: Physics of Medical Imaging
Ehsan Samei; Jiang Hsieh, Editor(s)
© SPIE. Terms of Use