The most important parameter that reflects the balance between oxygen supply and demand in tissues is the
mitochondrial NADH redox state that could be monitored In vivo. Nevertheless single parameter monitoring is limited
in the interpretation capacity of the very complicated pathophysiological events, therefore three more parameters were
added to the NADH and the multiparametric monitoring system was used in experimental and clinical studies.
In our previous paper1 we described the CritiView (CRV1) including a fiber optic probe that monitor four physiological
parameters in real time. In the new model (CRV3) several factors such as UV safety, size and price of the device were
improved significantly.
The CRV3 enable to monitor the various parameters in three different locations in the tissue thus increasing the
reliability of the data due to the better statistics. The connection between the device and the monitored tissue could be
done by various types of probes. The main probe that was tested also in clinical studies was a special 3 points probe that
includes 9 optical fibers (3 in each point) that was embedded in a three way Foley catheter. This catheter enabled the
monitoring of urethral wall vitality as an indicator of the development of body metabolic emergency state.
The three point probe was tested in the brain exposed to the lack of oxygen (Anoxia, Hypoxia or Ischemia). A decrease
in blood oxygenation and a large increase in mitochondrial NADH fluorescence were recorded. The microcirculatory
blood flow increased during anoxia and hypoxia and decreased significantly under ischemia.
Paper Details
Date Published: 15 February 2006
PDF: 10 pages
Proc. SPIE 6083, Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI, 60830Z (15 February 2006); doi: 10.1117/12.675745
Published in SPIE Proceedings Vol. 6083:
Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI
Israel Gannot, Editor(s)
PDF: 10 pages
Proc. SPIE 6083, Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI, 60830Z (15 February 2006); doi: 10.1117/12.675745
Show Author Affiliations
Avraham Mayevsky, Bar Ilan Univ. (Israel)
CritiSense Ltd. (Israel)
Yoram Blum, CritiSense Ltd. (Israel)
Nava Dekel, CritiSense Ltd. (Israel)
Assaf Deutsch, CritiSense Ltd. (Israel)
Rafael Halfon, CritiSense Ltd. (Israel)
CritiSense Ltd. (Israel)
Yoram Blum, CritiSense Ltd. (Israel)
Nava Dekel, CritiSense Ltd. (Israel)
Assaf Deutsch, CritiSense Ltd. (Israel)
Rafael Halfon, CritiSense Ltd. (Israel)
Shlomi Kremer, CritiSense Ltd. (Israel)
Eliyahu Pewzner, CritiSense Ltd. (Israel)
Efrat Sherman, Bar Ilan Univ. (Israel)
Ofer Barnea, Tel Aviv Univ. (Israel)
Eliyahu Pewzner, CritiSense Ltd. (Israel)
Efrat Sherman, Bar Ilan Univ. (Israel)
Ofer Barnea, Tel Aviv Univ. (Israel)
Published in SPIE Proceedings Vol. 6083:
Optical Fibers and Sensors for Medical Diagnostics and Treatment Applications VI
Israel Gannot, Editor(s)
© SPIE. Terms of Use