Calibration of the visible/near-infrared channels of the along-track scanning radiometer-2 (ATSR-2)
Author(s): D. L. Smith; Peter David Read; C. T. Mutlow
Format Member Price Non-Member Price
PDF $17.00 $21.00

Paper Abstract

An in-flight visible calibration system, VISCAL, is used to calibrate the visible/near infra-red channels of ATSR-2. The in-flight monitoring of the VISCAL is described and results presented. Data shows that the visible channels are affected by condensation effects as well as some long term degradation. The accuracy of the measured ground reflectances is dependent on the long term stability of the calibration. An area within the Libyan Desert has been used to determine calibration drift rates of 0.5% (1.6 micrometer), 1.8% (0.87 micrometer), 0.8% (0.66 micrometer) and 1.7% (0.56 micrometer) per year.

Paper Details

Date Published: 31 December 1997
PDF: 10 pages
Proc. SPIE 3221, Sensors, Systems, and Next-Generation Satellites, (31 December 1997); doi: 10.1117/12.298114