Share Email Print
cover

Proceedings Paper

Spot decomposition in a novel pencil beam scanning proton computed tomography
Author(s): J. Zhou; X. Li; P. Kabolizadeh; X. Yang; D. Yan; C. Stevens; T. Liu; X. Ding
Format Member Price Non-Member Price
PDF $17.00 $21.00

Paper Abstract

Range uncertainty is one of the most critical obstacles in proton therapy. Proton computed tomography (pCT) can potentially directly reconstruct relative stopping power (RSP) within 1%. We report the very first proton imaging technique based on spot decomposition within each spot for increasing efficiency of pCT using pencil beam scanning (PBS) technique. A 14cm-diameter cylinder water phantom was used in our simulation, embedded with three groups of cylinders (bone, muscle, and adipose, respectively). Each group of cylinders contains three different sizes (2cm, 1cm, and 3mm in diameter). A TOPAS Monte-Carlo model was developed simulating the on-board pCT image acquisition on a PBS gantry. A phase scorer was used to simulate a multi-layer pixelated proton residual energy detector. Each proton spot was divided into 19 overlapping sub-spots, and residual energy statistics were calculated for each sub-spot and were assigned to the nearest detector pixel. One hundred eighty projections were generated with 882 spots on each 10x10cm projection, using 200 MeV protons. Spot spacing and size (1-sigma) were 5.4mm and 6-8mm, respectively, on the detector. A direct path of proton transport route was assumed in an FDK-based reconstruction. pCT imaging dose was calculated, and the accuracy of RSP reconstruction was analyzed. Total dose to the phantom was 0.4mGy. The reconstructed mean RSP were 0.999(-0.1%), 0.969(-0.03%), 1.028(-0.2%), and 1.703(-0.9%) for the water, adipose, muscle, and bone, respectively. The STDs of the reconstructed RSP were <1.0%. This study has provided a new framework and offered an efficient approach for pCT acquisition and reconstruction.

Paper Details

Date Published: 1 March 2019
PDF: 8 pages
Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 109485Q (1 March 2019); doi: 10.1117/12.2512046
Show Author Affiliations
J. Zhou, Emory Univ. (United States)
Beaumont Hospital (United States)
X. Li, Beaumont Hospital (United States)
P. Kabolizadeh, Beaumont Hospital (United States)
X. Yang, Emory Univ. (United States)
D. Yan, Beaumont Hospital (United States)
C. Stevens, Beaumont Hospital (United States)
T. Liu, Emory Univ. (United States)
X. Ding, Beaumont Hospital (United States)


Published in SPIE Proceedings Vol. 10948:
Medical Imaging 2019: Physics of Medical Imaging
Taly Gilat Schmidt; Guang-Hong Chen; Hilde Bosmans, Editor(s)

© SPIE. Terms of Use
Back to Top