Improved wedge scatter correction for multi-slice CT system
Author(s): Yang Wang; Karl Stierstorfer; Martin Petersilka; Yi Tian
Format Member Price Non-Member Price
PDF $17.00 $21.00

Paper Abstract

In modern multi-slice CT systems, bowtie shape wedge filter is widely used for optimizing patient received radiation dose distribution. Since wedge filter is usually fixed in the path of X-ray, the induced scatter radiation could give impact on image quality under certain cases. In order to compensate this extra scatter radiation and improve the image quality, we introduced a wedge scatter correction algorithm integrated in the raw data pre-processing workflow. After the algorithm is implemented on our latest CT systems, the improvement can be seen from both water phantom images and clinical patient images.

Paper Details

Date Published: 1 March 2019
PDF: 7 pages
Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 109482B (1 March 2019); doi: 10.1117/12.2511910