Proceedings Paper
Design of precision hydrodynamic bearings for goniometer with radial gratingsFormat | Member Price | Non-Member Price |
---|---|---|
$17.00 | $21.00 |
Paper Abstract
In order to enhance the measuring accuracy of angular displacement, it is important to form high accuracy and superior quality of moire fringe signals. They are strongly affected by the radial disturbance of grating disk which is mounted on the shaft of bearing system. Specifications for the hydrodynamic bearing system often call for much higher radial accuracy and stability. This requirement poses unique challenges for the design of the bearing system. In this paper, the design of precision three-lobe hydrodynamic bearings is described, and lubricant performance of some structure parameters is analyzed. The radial accuracy of this bearing system is within 0.023 micrometers .
Paper Details
Date Published: 6 September 1995
PDF: 9 pages
Proc. SPIE 2542, Optomechanical and Precision Instrument Design, (6 September 1995); doi: 10.1117/12.218679
Published in SPIE Proceedings Vol. 2542:
Optomechanical and Precision Instrument Design
Alson E. Hatheway, Editor(s)
PDF: 9 pages
Proc. SPIE 2542, Optomechanical and Precision Instrument Design, (6 September 1995); doi: 10.1117/12.218679
Show Author Affiliations
Jianshe Wang, Changchun Institute of Optics and Fine Mechanics (China)
Yingshi Zhu, Changchun Institute of Optics and Fine Mechanics (China)
Published in SPIE Proceedings Vol. 2542:
Optomechanical and Precision Instrument Design
Alson E. Hatheway, Editor(s)
© SPIE. Terms of Use