Comparison between angle interferometer and angle encoder during calibration of autocollimatorFormat | Member Price | Non-Member Price |
---|---|---|
$17.00 | $21.00 |
Paper Abstract
An angle interferometer was set up using concept ‘ratio of two lengths’ and an angle encoder was set up using concept ‘subdivision of full circle (2π rad=360°)’ at the National Institute of Metrology, China (NIM). For the analysis of the systematic errors of each device, two autocollimator calibration systems were separately set up based on the angle interferometer and the angle encoder with a similar measuring uncertainty (around 0.1″). An autocollimator was calibrated using two systems in the same measurement range (±1000″) and the same measurement step (10″). The systematic errors of each system were found through comparison between their original calibration results. The compensation curves were calculated using the analysis results, and two systems’ original calibration results were compensated according to two systems’ compensation curves. The maximum difference between the compensated calibration results of two systems was 0.05″ which is lower than measuring uncertainty of each system.
Paper Details
Date Published: 6 March 2015
PDF: 11 pages
Proc. SPIE 9446, Ninth International Symposium on Precision Engineering Measurement and Instrumentation, 944624 (6 March 2015); doi: 10.1117/12.2180902
Published in SPIE Proceedings Vol. 9446:
Ninth International Symposium on Precision Engineering Measurement and Instrumentation
Junning Cui; Jiubin Tan; Xianfang Wen, Editor(s)
PDF: 11 pages
Proc. SPIE 9446, Ninth International Symposium on Precision Engineering Measurement and Instrumentation, 944624 (6 March 2015); doi: 10.1117/12.2180902
Show Author Affiliations
He Yan Wang, National Institute of Metrology (China)