Proceedings Paper
Circular tomosynthesis for neuro perfusion imaging on an interventional C-armFormat | Member Price | Non-Member Price |
---|---|---|
$17.00 | $21.00 |
Paper Abstract
There is a clinical need to improve cerebral perfusion assessment during the treatment of ischemic stroke in the interventional suite. The clinician is able to determine whether the arterial blockage was successfully opened but is unable to sufficiently assess blood flow through the parenchyma. C-arm spin acquisitions can image the cerebral blood volume (CBV) but are challenged to capture the temporal dynamics of the iodinated contrast bolus, which is required to derive, e.g., cerebral blood flow (CBF) and mean transit time (MTT). Here we propose to utilize a circular tomosynthesis acquisition on the C-arm to achieve the necessary temporal sampling of the volume at the cost of incomplete data. We address the incomplete data problem by using tools from compressed sensing and incorporate temporal interpolation to improve our temporal resolution. A CT neuro perfusion data set is utilized for generating a dynamic (4D) volumetric model from which simulated tomo projections are generated. The 4D model is also used as a ground truth reference for performance evaluation. The performance that may be achieved with the tomo acquisition and 4D reconstruction (under simulation conditions, i.e., without considering data fidelity limitations due to imaging physics and imaging chain) is evaluated. In the considered scenario, good agreement between the ground truth and the tomo reconstruction in the parenchyma was achieved.
Paper Details
Date Published: 18 March 2015
PDF: 9 pages
Proc. SPIE 9412, Medical Imaging 2015: Physics of Medical Imaging, 94122A (18 March 2015); doi: 10.1117/12.2082695
Published in SPIE Proceedings Vol. 9412:
Medical Imaging 2015: Physics of Medical Imaging
Christoph Hoeschen; Despina Kontos, Editor(s)
PDF: 9 pages
Proc. SPIE 9412, Medical Imaging 2015: Physics of Medical Imaging, 94122A (18 March 2015); doi: 10.1117/12.2082695
Show Author Affiliations
Bernhard E. Claus, GE Global Research Ctr. (United States)
David A. Langan, GE Global Research Ctr. (United States)
David A. Langan, GE Global Research Ctr. (United States)
Omar Al Assad, GE Global Research Ctr. (United States)
Xin Wang, GE Global Research Ctr. (United States)
Xin Wang, GE Global Research Ctr. (United States)
Published in SPIE Proceedings Vol. 9412:
Medical Imaging 2015: Physics of Medical Imaging
Christoph Hoeschen; Despina Kontos, Editor(s)
© SPIE. Terms of Use