Detection of low-dose CT reconstruction artifacts using a bi-modal approachFormat | Member Price | Non-Member Price |
---|---|---|
$17.00 | $21.00 |
Paper Abstract
Low-dose Computed Tomography (CT) has the benefit of exposing patients to less radiation. However,
low dose CT requires special reconstruction techniques to improve the clarity of the image. Unfortunately,
these special reconstruction techniques often cannot remove all of the low-dose artifacts. It is important to
recognize these artifacts else we run the risk of obscuring important detail or adding false features. In this
work, we present a simple scheme which allows us to detect these artifacts. Our technique applies to the
specific low-dose CT strategy in which the number of X-ray views taken from the patient is reduced. The
first step uses directional interpolation in the low dose sinogram to add more views. While the image
created from this interpolated sinogram does not have any artifacts it lacks significantly in clarity due to
blurring. Our scheme then compares this image with the image created directly with a low-dose CT
reconstruction technique which has better detail but also some remaining artifacts. The comparison
reveals these artifacts which we then remove by simple pixel replacement.
Paper Details
Date Published: 6 March 2013
PDF: 7 pages
Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 86683M (6 March 2013); doi: 10.1117/12.2008170
Published in SPIE Proceedings Vol. 8668:
Medical Imaging 2013: Physics of Medical Imaging
Robert M. Nishikawa; Bruce R. Whiting; Christoph Hoeschen, Editor(s)
PDF: 7 pages
Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 86683M (6 March 2013); doi: 10.1117/12.2008170
Show Author Affiliations
Published in SPIE Proceedings Vol. 8668:
Medical Imaging 2013: Physics of Medical Imaging
Robert M. Nishikawa; Bruce R. Whiting; Christoph Hoeschen, Editor(s)
© SPIE. Terms of Use