Share Email Print
cover

Proceedings Paper

A comparison study of sinogram- and image-domain penalized re-weighted least-squares approaches to noise reduction for low-dose cone-beam CT
Author(s): Hao Zhang; Yan Liu; Hao Han; Jing Wang; Jianhua Ma; Lihong Li; Zhengrong Liang
Format Member Price Non-Member Price
PDF $17.00 $21.00

Paper Abstract

Reducing X-ray exposure to the patients is one of the major research efforts in the computed tomography (CT) field, and one of the common strategies to achieve it is to lower the mAs setting (by lowering the X-ray tube current and/or shortening the exposure time) in currently available CT scanners. However, the image quality from low mAs acquisition is severely degraded due to excessive quantum noise, if no adequate noise control is applied during image reconstruction. Different from filter-based algorithms, statistical reconstruction algorithms model the statistical property of the noise using a cost function and minimize the cost function for an optimal solution in statistical sense. The algorithms have shown to be feasible and effective in both sinogram and image domain. In our previous researches, we proposed penalized reweighted least-squares (PRWLS) approaches to sinogram noise reduction and image reconstruction for low-dose CT imaging, which are in this statistical category. This work is a continuation of the research along this direction and aims to compare the reconstruction quality of two different PRWLS implementations for low-dose cone-beam CT reconstruction: (1) PRWLS sinogram restoration followed by analytical Feldkamp-Davis- Kress reconstruction, (2) fully iterative PRWLS image reconstruction. Inspired by our recent study on the variance of low-mAs projection data in presence of electric noise background, a more accurate weight was adopted in the weighted least-squares term. An anisotropic quadratic form penalty was utilized in both PRWLS implementations to preserve edges during noise reduction. Experiments using the CatPhan® 600 phantom and anthropomorphic head phantom were carried to study the relevant performance of these two implementations on image reconstruction. The results revealed that the implementation (2) can outperform implementation (1) in terms of noise-resolution tradeoff measurement and analysis of the reconstructed small objects due to its matched image edge-preserved penalty in the image domain. However, those gains are offset by the cost of increased computational time. Thus, further examination of real patient data is necessary to show the clinical significance of the iterative PRWLS image reconstruction over the PRWLS sinogram restoration.

Paper Details

Date Published: 6 March 2013
PDF: 8 pages
Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 86683E (6 March 2013); doi: 10.1117/12.2007111
Show Author Affiliations
Hao Zhang, Stony Brook Univ., SUNY (United States)
Yan Liu, Stony Brook Univ., SUNY (United States)
Hao Han, Stony Brook Univ., SUNY (United States)
Jing Wang, Univ. of Texas Southwestern Medical Ctr. (United States)
Jianhua Ma, Stony Brook Univ., SUNY (United States)
Southern Medical Univ. (China)
Lihong Li, City Univ. of New York, SUNY (United States)
Zhengrong Liang, Stony Brook Univ., SUNY (United States)


Published in SPIE Proceedings Vol. 8668:
Medical Imaging 2013: Physics of Medical Imaging
Robert M. Nishikawa; Bruce R. Whiting; Christoph Hoeschen, Editor(s)

© SPIE. Terms of Use