Proceedings Paper
A tissue mimicking phantom model for applications combining light and ultrasoundFormat | Member Price | Non-Member Price |
---|---|---|
$17.00 | $21.00 |
Paper Abstract
We describe a stable and reproducible liquid tissue mimicking phantom optimized for applications involving both
ultrasound and light waves. The phantom has optical and acoustic properties similar to soft biological tissue. The base
material is Glycerol. The TiO2 is added to the Glycerol as scattering particles. An absorbing dye is added to obtain
desired absorptions in the Near IR range. The phantom's optical absorption was measured by Spatially Resolved
Spectroscopy (SRS). In addition, the optical properties were calculated based on the spatial decay of an acousto-optic
signal generated in the phantom, and were compared to those obtained with SRS.
Paper Details
Date Published: 28 February 2013
PDF: 7 pages
Proc. SPIE 8583, Design and Performance Validation of Phantoms Used in Conjunction with Optical Measurement of Tissue V, 858307 (28 February 2013); doi: 10.1117/12.2003528
Published in SPIE Proceedings Vol. 8583:
Design and Performance Validation of Phantoms Used in Conjunction with Optical Measurement of Tissue V
Robert J. Nordstrom, Editor(s)
PDF: 7 pages
Proc. SPIE 8583, Design and Performance Validation of Phantoms Used in Conjunction with Optical Measurement of Tissue V, 858307 (28 February 2013); doi: 10.1117/12.2003528
Show Author Affiliations
A. Ron, Ornim Medical Ltd. (Israel)
N. Racheli, Ornim Medical Ltd. (Israel)
N. Racheli, Ornim Medical Ltd. (Israel)
Published in SPIE Proceedings Vol. 8583:
Design and Performance Validation of Phantoms Used in Conjunction with Optical Measurement of Tissue V
Robert J. Nordstrom, Editor(s)
© SPIE. Terms of Use