The design and operation of a gas millimeter-wave spectrometer based on a tunable generator orotron are presented. The absorption cell occupies a part of the high quality (Q approximately equals 104) orotron Fabry-Perot cavity. The limit of sensitivity (gamma) min approximately equals (3- 5)X 10-10 cm-1 has been achieved with source frequency modulation at the output band of a receiver of 1 Hz. The narrow spectral width of the orotron radiation (10 - 15 kHz without locking) provides the Doppler resolution of spectral lines.

Paper Details

Date Published: 26 January 1994
PDF: 5 pages
Proc. SPIE 2205, 11th Symposium and School on High-Resolution Molecular Spectroscopy, (26 January 1994); doi: 10.1117/12.166263
Show Author Affiliations
B. S. Dumesh, Institute of Spectroscopy (Russia)
V. D. Gorbatenkov, Institute of Spectroscopy (Russia)
V. P. Kostromin, Institute of Spectroscopy (Russia)
F. S. Rusin, Institute of Spectroscopy (Russia)
L. A. Surin, Institute of Spectroscopy (Russia)