Fabry-Perot Interferometer


Excerpt from Field Guide to Spectroscopy

A Fabry-Perot interferometer is very similar to a Fabry-Perot filter. However, instead of a thin layer of some medium sandwiched between two mirrors, the Fabry-Perot interferometer has two parallel, highly reflective mirrors separated by an air gap several millimeters to several centimeters in width.

Light of wavelength λ constructively interferes when the following relationship is satisfied:

mλ = 2d cosθ

where m is the order and θ is the angle of the incoming light.

Fabry-Perot interferometer

If one of the mirrors is mechanically moved back and forth (most systems use a piezoelectric crystal), the wavelength that constructively interferes at a particular angle θ varies, and the system acts as an interferometer. Since the angle of refraction also depends on the index of refraction of the gap between the mirrors, Fabry-Perot interferometers can also be used with constant d but by changing the pressure of some filler gas.

Fabry-Perot interferometers are capable of very high resolution, able to distinguish spectral lines less than a nanometer apart.
Citation:

D. W. Ball, Field Guide to Spectroscopy, SPIE Press, Bellingham, WA (2006).