How to effectively evaluate a direct laser lithography system

Measurement uncertainty quantifies performance-related doubt and evaluates the reliability of a direct laser lithography system.
20 January 2012
Dong-Ik Kim, Geon Hee Kim and Hyug-Gyo Rhee

It is common knowledge that obtaining high-quality fabricated products relies on the performance of the instruments used to make them. Accordingly, fully evaluating the reliability of the instruments requires taking into account the ‘measurement uncertainty’ of the results.1, 2 Measurements convey information about a property of a thing. For example, a measurement might indicate the weight or length of an object. It gives a number to the property of interest. Well-made rulers, clocks, and thermometers might be assumed to be trustworthy and to give the right answers. But for every measurement—even the most careful—there is always a margin of doubt. Quantifying the degree of this uncertainty is key to increasing reliability. The method is complicated by the measuring instrument, the object being measured, operator skill, sampling issues, and the environment, and, as a result, it is not yet widely used in a field as specialized as lithography.

To illustrate the utility of measurement uncertainty, we applied it to a direct laser lithography system (DLLS) that we developed. Its working size is up to 360mm, and its minimum line width is about 0.5μm, enabling it to fabricate various computer-generated holograms (CGHs) from 5 to 360mm in diameter.3 To assure the performance of the DLLS, we evaluated the moving stages and the writing head module.

For the moving stages, evaluation consisted of the standard uncertainty of a line-positioning error in the radial direction. This error, defined as Δ, is due mainly to the linearity (Δl=p2−p1) and the angular motion error Δa of a linear stage, the run-out error Δr of a rotary stage, and the centering error Δc of the DLLS (see Figure 1).


Figure 1. Schematic of the direct laser lithography system positioning error. (a) Linearity error Δland centering error Δc. (b) Angular motion error Δa. L: Distance from the connecting part of the writing head and the moving stage to the focal point.

First, we measured Δl using a commercial heterodyne interferometer on the moving axis of the linear stage. The uncertainty of Δl, defined as ul, was 4.0nm based on 30 repeated measurements. The second error, due to angular motion of the linear stage coupled with a distance factor (the distance from the connecting part of the writing head and the moving stage to the focal point), is directly measured by an autocollimator and a commercial coordinate-measuring instrument (see Figure 2).


Figure 2. (a) Linearity error and (b) angular motion error of the linear stage.

The uncertainty of the angular error ua was 8.3nm within a 180mm range. For the run-out error of the rotary stage, we fabricated a 0.8mm circle. We measured the circle with a commercial white-light scanning interferometer (see Figure 3). Using an edge-detection algorithm4 and a circle–fitting method, we estimated Δr. The uncertainty ur of Δr was 2.5nm at 600rpm. To estimate Δc (another dominant factor), we fabricated a spiral pattern on a substrate and extracted Δc using a line-fitting algorithm. In our case, Δc and its uncertainty, uc, were 12 and 6.0nm, respectively. A combined uΔ, which is the square root of the sum of each uncertainty squared, was 11.3nm. We obtained different uncertainties from 30 repeated measurements for better estimation.

The overall (so-called expanded) uncertainty, stated at level of confidence of 95%, can be determined using a coverage factor, k. Multiplying the combined standard uncertainty uΔ by the coverage factor gives the expanded uncertainty, defined as U=kuΔ. The coverage factor used in this case was k=2. The expanded uncertainty U was 22.6nm. The calculated wavefront error due to U was about 5.19nm for a typical case of CGH. We measured the wavefront error in a 98mm-diameter CGH using a commercial Fizeau interferometer. The result was 0.0063λ (λ=632:8nm) in rms value (see Figure 4).


Figure 3. (a) Circle pattern and (b) run-out error Δrof the linear stage.

Figure 4. (a) Photographic view and (b) wavefront error of the 98mm-diameter computer-generated hologram. (c) The wavefront error after application of the absolute testing scheme.

Finally, we estimated the focusing error of the writing head module. The calculated depth of focus with respect to the writing beam of 488nm and the numerical aperture of 0.9 was 301.2nm. The autofocus controller showed a 280nm focusing error. This result is sufficient to satisfy the tolerance specification. Ultimately, we successfully fabricated a 300mm-diameter CGH using the results reported here.

In summary, a measurement result can be only complete if it is accompanied by a statement of the uncertainty of the process. Consequently, applying measurement reliability to the DLLS enabled us to objectively estimate uncertainties for the first time. As a next step, we plan to comprehensively evaluate the DLLS, including the uncertainty of the stability of the light source.


Dong-Ik Kim, Geon Hee Kim
Division of Instrument Development
Korea Basic Science Institute
Daejeon, South Korea
Hyug-Gyo Rhee
Center for Space Optics
Korea Research Institute of Standards and Science
Daejeon, South Korea

References:
1. http://www.nist.gov/pml/pubs/tn1297/index.cfm Guidelines for evaluating and expressing the uncertainty of NIST measurement results. Accessed 5 January 2012.
2. S. Bell, A Beginner's Guide to Uncertainty of Measurement, National Physical Laboratory, 2001.
3. D.-I. Kim, H.-G. Rhee, G.-H. Kim, Performance evaluation of direct laser lithographic system for rotationally symmetric diffractive optical elements, Proc. SPIE 8249, 2012. In press.
4. J. Jin, H. Rhee, S. Kim, Metrological atomic force microscopy integrated with a modified two-point diffraction interferometer, Meas. Sci. Technol 20, pp. 105302, 2009. doi:10.1088/0957-0233/20/10/105302
Recent News
PREMIUM CONTENT
Sign in to read the full article
Create a free SPIE account to get access to
premium articles and original research