Described are the comparative testing of two type of electron-optical recording instruments, namely, TDS simplified streak device designed and manufactured at the Centre of Scientific Instruments, GDR Academy of Sciences, and anglo-soviet Imacon 500 picosecond streak camera fitted with PV-001 tube. The Imacon 500 streak camera was matched with a SIT vidicon readout system. Temporal resolution, dynamic range and input sensitivity were estimated for the same recording conditions. It was shown that for 6 ps laser probe pulses the TDS dynamic sensitivity at 530 nm input radiation is only four times less comparative to the commercial Imacon 500 system, while both systems have almost the same dynamic range.

Paper Details

Date Published: 7 June 1989
PDF: 4 pages
Proc. SPIE 1032, 18th Intl Congress on High Speed Photography and Photonics, (7 June 1989); doi: 10.1117/12.969078
Show Author Affiliations
A. V. Babushkin, USSR Academy of Sciences (USSR)
E. Jaeger, GDR Academy of Sciences (GDR)
V. E. Postovalov, USSR Academy of Sciences (USSR)
C. Rempel, GDR Academy of Sciences (GDR)
M. Ya. Schelev, USSR Academy of Sciences (USSR)