Paper Abstract

The operating principle and design of synchronous streak camera systems are reviewed and experimental results are included to show that a temporal resolution ~1. ps has been demonstrated. A selection of applications in photophysics and photochemistry are briefly described to illustrate some of the unique features of this diagnostic technique. The further developments relating to an extended range of spectral sensitivity and femtosecond resolution are also briefly discussed.

Paper Details

Date Published: 1 March 1983
PDF: 12 pages
Proc. SPIE 0348, 15th Intl Congress on High Speed Photography and Photonics, (1 March 1983); doi: 10.1117/12.967699