We present new results in four areas of microchannel plate (MCP) X-ray detector operation. The performance in pulse counting mode of MCPs with 8 micron channel diameters is reported. The effects on MCP quantum detection efficiency and energy discrimination of multiple CsI coatings are described. A new mode of operation of two-stage multipliers is evaluated. Replacing the conventional electron-accelerating inter-plate potential difference by a retarding field is shown to result in definite advantages with regard to X-ray energy discrimination and detector lifetime. The source of the MCP internal background is discussed.