X-Ray Calibration Of The Rosat Position Sensitive Proportional Counter II. The Energy Calibration
Author(s): U. G. Briel; W. Burkert; E. Pfeffermann
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

The ROSAT position sensitive proportional counter is a multi wire X-ray detector which uses a mixture of Argon, Xenon and Methan as counter gas. In the X-ray energy band from 0.1 keV to 2.2 keV, this gas shows several absorption edges, caused by Argon-L and Xenon-N shells between 100 eV and 350 eV, and Xenon-M shells between 670 eV and 1150 eV. To measure the anode pulse height as a function of X-ray energy, we set up an X-ray spectrometer, using a transmission grating with 1000 lines per mm and a high intensity X-ray continuum source. The spectrometer had a resolution of 0.3 A. Besides the well known discontinuity of the pulse height at the Argon-L III/II edges, we could identify almost all Xenon edges in our energy range. Between the edges, the anode pulse height is proportional to the X-ray energy. At the edges, the pulse height drops, corresponding to an energy loss of a few electron volt. We will describe the measurements and present the results.

Paper Details

Date Published: 27 November 1989
PDF: 7 pages
Proc. SPIE 1159, EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy and Atomic Physics, (27 November 1989); doi: 10.1117/12.962584
Show Author Affiliations
U. G. Briel, Max-Planck-Institut fur Physik und Astrophysik (Germany)
W. Burkert, Max-Planck-Institut fur Physik und Astrophysik (Germany)
E. Pfeffermann, Max-Planck-Institut fur Physik und Astrophysik (Germany)


Published in SPIE Proceedings Vol. 1159:
EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy and Atomic Physics
Charles J. Hailey; Oswald H. W. Siegmund, Editor(s)

© SPIE. Terms of Use
Back to Top