Positioning Error Induced By Axial Sampling In Positron Emission Tomography
Author(s): Bernard Bendriem; Stephen L. Dewey; David J. Schlyer
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

A multi-slice Positron Camera organizes data as a set of parallel slices whose thickness is defined by the axial response function (ARF) of the tomograph. This ARF is usually not uniform and the sensitivity is maximum for a point source located at the center of the slice. As a consequence, the partial volume effect (quantitation loss for structures that are smaller than two to three times the FWHM resolution) will he dependent not only on the size of the structure but also on its relative position with regard to the ARF. We have developed a computer simulation to analyze this effect. Objects of different size and shape are examined and it is shown that on a system with a 6 mm axial resolution (FWHM), a shift of 3 mm can introduce more than 10% variation on the final measurement. The simulation is confirmed by experimental results on a high resolution PET. We conclude that positioning and repositioning must be handled with great care in order to make reproducible measurements and allow accurate partial volume corrections.

Paper Details

Date Published: 18 October 1989
PDF: 6 pages
Proc. SPIE 1137, Science and Engineering of Medical Imaging, (18 October 1989); doi: 10.1117/12.961719
Show Author Affiliations
Bernard Bendriem, Brookhaven National Laboratory (United States)
Stephen L. Dewey, Brookhaven National Laboratory (United States)
David J. Schlyer, Brookhaven National Laboratory (United States)


Published in SPIE Proceedings Vol. 1137:
Science and Engineering of Medical Imaging

© SPIE. Terms of Use
Back to Top