Share Email Print
cover

Proceedings Paper

Unique Optical System For Conversion Of Existing Photometers Into Scanning Microphotometers
Author(s): Thomas H. Bulpitt; Richard A. Walker
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

There are many thousands of luminance photometers in the field - both telephotometers and microphotometers; however very few of these have scanning microphotometric capability. This paper describes a unique optical system that can be field-interchanged with the standard objective lens, without requiring any modification of the photometer's internal optical system, to convert an existing photometer into a high-resolution scanning microphotometer. One embodiment, called the "MicroScannerTM Spatial Scanner", will be described in detail. Used with any standard Pritchard Photometer, this scanner provides variable-speed scanning of a half-inch (12 mm) of object space with field coverage as fine as 0.0001-inches (0.0025 mm) or as broad as 0.39-inches (10 mm). The system features photometric uniformity within ± 0 5% over the entire scan distance, freedom from polarization error, and a half-inch scan distance independent of choice objective lens.

Paper Details

Date Published: 15 November 1979
PDF: 9 pages
Proc. SPIE 0196, Measurements of Optical Radiations, (15 November 1979); doi: 10.1117/12.957964
Show Author Affiliations
Thomas H. Bulpitt, Kollmorgen Corporation (United States)
Richard A. Walker, Kollmorgen Corporation (United States)


Published in SPIE Proceedings Vol. 0196:
Measurements of Optical Radiations
Harold P. Field; Edward F. Zalewski; Frederic M. Zweibaum, Editor(s)

© SPIE. Terms of Use
Back to Top