Evaluation Of Micro-Focal Spot X-Ray Tubes And Rare Earth Oxysulfide Intensifying Screens In Magnification Radiography
Author(s): H. Eisenberg; M. Braun; B. Arnold; W. Holland; R. Gould
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

The results of some preliminary measurements of the modulation transfer functions of (1) micro-focal spots of prototype x-ray tubes and (2) rare-earth oxysulfide intensifying screens are presented. The micro-focus tubes when used in conjunction with the rare earth screen-film systems have sufficient output to stop biologic motion in most clinical examinations. Optimum magnification is arrived at by use of the "effective sampling aperture" as suggested by Wagner (Ref. 1). An optimized micro-focus-rare-earth system is then compared to a conventional system of similar speed using calcium tungstate screens and 0.3 mm focal spot. The results show that a considerable improvement in contrast transfer is obtainable with the new system. Supportative radiolpgical evidence is presented.

Paper Details

Date Published: 1 May 1974
PDF: 8 pages
Proc. SPIE 0056, Medical X-Ray Photo-Optical Systems Evaluation, (1 May 1974); doi: 10.1117/12.954288
Show Author Affiliations
H. Eisenberg, Beth Israel Hospital (United States)
M. Braun, Machlett Laboratories (United States)
B. Arnold, Beth Israel Hospital (United States)
W. Holland, Machlett Laboratories (United States)
R. Gould, Beth Israel Hospital (United States)

Published in SPIE Proceedings Vol. 0056:
Medical X-Ray Photo-Optical Systems Evaluation
David J. Goodenough; Robert F. Wagner; Kenneth E. Weaver, Editor(s)