In attempting to specify the performance of an x-ray tube focal spot, a number of factors must be taken into consideration which influence the spatial and intensity distribution of the focal spot: Tube potential, tube current, off-focus radiation, and off-axis variations. In addition, a knowledge of the tube loading capabilities is necessary to adequately predict the focal spot's role in the total imaging process. This paper presents techniques that were developed to investigate some of these factors and some typical results of work to date.

Paper Details

Date Published: 1 May 1974
PDF: 9 pages
Proc. SPIE 0056, Medical X-Ray Photo-Optical Systems Evaluation, (1 May 1974); doi: 10.1117/12.954284