Scanned Projection Radiography Using A Time-Delay-And-Integrate Scanning CCD
Author(s): Steven L. Fritz
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

A scanned projection radiography sensor has been developed using a time-delay-and-integrate imaging CCD optically coupled to an X-ray intensifying screen. A mathematical model has been developed to predict the efficiency of information transfer between intensifying screen and CCD. This model predicts that for an 8-inch field of view the CCD should capture one electron for each X-ray photon absorbed in the screen. Variance in the number of electrons captured should reduce the DQE of the screen by a factor of 0.7. This almost exactly offsets the effect of scattered radiation in film/screen systems, suggesting that the new detector could have image quality equivalent to a screen/film/grid system. Experimental data shows that the model correctly predicts the light photon transfer efficiency of the system but that the image noise is dominated by switching noise from the CCD. With a low-noise camera design the system should be equal or superior to a screen/film/grid system for a wide range of practical imaging situations.

Paper Details

Date Published: 1 May 1989
PDF: 10 pages
Proc. SPIE 1090, Medical Imaging III: Image Formation, (1 May 1989); doi: 10.1117/12.953218