Automatic Determination Of Image Quality Parameters In Digital Radiographic Imaging Systems
Author(s): Martin Brok; Cornelis H. Slump
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

It is shown, that the Image Quality parameters, Modulation Transfer Function (MTF) and noise can be evaluated digitally with high accuracy. The systems MTF is measured by the use of a 5 slits phantom. Simulation studies are showing the optimal phantom dimension and indicate the expected precision of the measurement. By comparison studies with conventional techniques the MTF measurement procedure is validated for the effects of pixel coarseness, noise and limiting resolution. An analytical expression for signal-to-noise ratio in a digital radiographic system is developed. Predicted values are compared with measured data.