Share Email Print
cover

Proceedings Paper

In Vitro Atherosclerosis: Correlation Between IV Ultrasonic Imaging And High Resolution MRI
Author(s): C. R. Meyer; E. H. Chiang; D. M. Williams; A. J. Buda
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Previous in vitro ultrasonic imaging in our laboratory with a prototype subassembly suitable for use in an imaging catheter was performed at 20 MHz using a rather large transducer having a 2 mm aperture focussed at 10 mm [1,2,3]. More recently ultrasonic images obtained at 30 MHz with a 1 mm active aperture focussed at 3 mm, have been compared with high resolution magnetic resonance images (MRI). Axial and lateral ultrasonic resolutions have improved from 100 and 450 μm, respectively, to 100 and 150 μm using the newer, smaller transducer. Signal-to-noise ratios as visualized in the resulting images remains unchanged, and penetration of normal vessel walls is complete.

Paper Details

Date Published: 8 August 1989
PDF: 3 pages
Proc. SPIE 1068, Catheter-Based Sensing and Imaging Technology, (8 August 1989); doi: 10.1117/12.952182
Show Author Affiliations
C. R. Meyer, University of Michigan Hospitals (United States)
E. H. Chiang, University of Michigan Hospitals (United States)
D. M. Williams, University of Michigan Hospitals (United States)
A. J. Buda, University of Michigan Hospitals (United States)


Published in SPIE Proceedings Vol. 1068: