Share Email Print
cover

Proceedings Paper

Ablation Of Cataractous Lens Tissue With Carbon Dioxide Laser Radiation
Author(s): L. M. Kukreja
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Ablation of cataractous human eye lens tissue with carbon dioxiqe laser beam of wave-length 10.6 μm is investigated at power densities upto 20 KW/cm2 with interaction time in the rnge of 4-20 msec. It is found that the ablation initiates at a fluence of about 26 J/cm2 and thereafter the ablated volume of the tissue increases with the laser fluence with a decreasing slope. A well defined zone of coagulation necrosis borders the laser ablated region and its width increases with the power density and interaction time of the laser beam. From the thermal behaviour of the cataractous lens tissue, it is found that at the onset of ablation, tissue attains a temperature of about 270°C. To confine this temperature to a reasonably small volume, short duration laser pulses will be ideal for ablating the cataractous lens.

Paper Details

Date Published: 11 September 1989
PDF: 5 pages
Proc. SPIE 1066, Laser Surgery: Advanced Characterization, Therapeutics, and Systems, (11 September 1989); doi: 10.1117/12.952022
Show Author Affiliations
L. M. Kukreja, Bhabha Atomic Research Centre (India)


Published in SPIE Proceedings Vol. 1066:
Laser Surgery: Advanced Characterization, Therapeutics, and Systems
Kazuhiko Atsumi; Norman R. Goldblatt; Stephen N. Joffe, Editor(s)

© SPIE. Terms of Use