Share Email Print

Proceedings Paper

A Radiometric Method Of Calibrating Infrared Sources, Using The Sr 5000 Infrared Spectroradiometer
Author(s): A. Daniels; D. Cabib; Z. Kopolovich; R A Buckwald
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Commercial infrared sources (commonly known as "blackbodies") are useful as reference sources in as much as they emit known amounts of radiation. These sources are usually calibrated by contact temperature sensors (thermocouples), or by a radiometric nulling method. The former method assumes that the radiator emissivity is equal to one; the latter is very time consuming. In this paper we present the following: i)spectroradiometric measurements of a blackbody, which show that the temperature predicted radiometrically agrees with the nulling method calibration within the blackbody calibration accuracy (±2C); ii) analysis of the errors present in the radiometric calibration, which show that an accuracy better than ±2C can be expected from the SR 5000: the SR 5000 is used with an effective temperature software package and an NBS traceable reference blackbody. The analysis can be used to predict the accuracy that can be reached with other radiometers, once their performance parameters are known.

Paper Details

Date Published: 7 December 1988
PDF: 10 pages
Proc. SPIE 0972, Infrared Technology XIV, (7 December 1988); doi: 10.1117/12.948307
Show Author Affiliations
A. Daniels, CI Systems (Israel), Ltd. (Israel)
D. Cabib, CI Systems (Israel), Ltd. (Israel)
Z. Kopolovich, CI Systems (Israel), Ltd. (Israel)
R A Buckwald, CI Systems (Israel), Ltd. (Israel)

Published in SPIE Proceedings Vol. 0972:
Infrared Technology XIV
Irving J. Spiro, Editor(s)

© SPIE. Terms of Use
Back to Top