Effects Of kVp On DSA Images
Author(s): S P Wang; S A Schoolman
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

This work reports the measured dependence of the DSA image's iodine contrast on kVp, patient thickness, X-ray scatter, and DSA system's veiling glare. It also reports the measured patient skin dose at a constant X-ray image intensifier (XRII) entrance dose for the various kVp and the patient thickness settings. It is found that, for any given patient skin dose, the DSA image quality is most optimal in the 80 - 100 kVp range. However, the DSA image quality is markedly reduced by increasing patient thickness and by the combined effect of X-ray scatter and the DSA system's veiling glare. Clinical examples indeed show that 80 - 100 kVp operation yields very satisfactory images for all range of patient thickness and for both intravenous and intra-arterial DSA examinations.

Paper Details

Date Published: 15 June 1984
PDF: 6 pages
Proc. SPIE 0454, Application of Optical Instrumentation in Medicine XII, (15 June 1984); doi: 10.1117/12.939343