Absolute Sensitivity Calibration from 20 to 430 of a Grazing Incidence Spectrometer with a Multi-element Spectral Detector
Author(s): J. L. Terry; H. L. Manning; E. S. Marmar
Format Member Price Non-Member Price
PDF $17.00 $21.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Two methods which together allow sensitivity calibration from 20 Å to 430 Å are described in detail. The first method, useful up to 120 Å, uses a low power source to generate K∝ X-rays which are alternately viewed by an absolute detector (a proportional counter) and the spectrometer. The second method extends that calibration to 430 Å. It relies on the 2:1 brightness ratio of bright doublet lines from impurity ions which have a single outer shell electron and which are present in hot, magnetically confined plasmas. It requires that the absolute sensitivity of the spectrometer be known at one wavelength point, and in practice requires a multi-element spectral detector.

Paper Details

Date Published: 12 August 1986
PDF: 7 pages
Proc. SPIE 0689, X-Ray Calibration: Techniques, Sources, and Detectors, (12 August 1986); doi: 10.1117/12.936564
Show Author Affiliations
J. L. Terry, M.I.T. (United States)
H. L. Manning, M.I.T. (United States)
E. S. Marmar, M.I.T. (United States)


Published in SPIE Proceedings Vol. 0689:
X-Ray Calibration: Techniques, Sources, and Detectors
Ping Lee; Paul D. Rockett, Editor(s)