Share Email Print
cover

Proceedings Paper

A Penning Discharge Source For Extreme Ultraviolet Calibration
Author(s): David S. Finley; Patrick Jelinsky; Stuart Bowyer; Roger F. Malina
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

We have developed a Penning discharge lamp for use in the calibration of instruments and components for the extreme ultraviolet. This source is sufficiently light and compact to make it suitable for mounting on the movable slit assembly of a grazing incidence Rowland circle monochromator. Because this is a continuous discharge source, it is suitable for use with photon counting detectors. Line radiation is provided both by the gas and by atoms sputtered off the interchangeable metal cathodes. Usable lines are produced by species as highly ionized as Ne IV and Al V. The wavelength coverage provided is such that a good density of emission lines is available down to wavelengths as short as 100Å. This source fills the gap between 100 and 300Å which is inadequately covered by the other available compact continuous radiation sources.

Paper Details

Date Published: 12 August 1986
PDF: 5 pages
Proc. SPIE 0689, X-Ray Calibration: Techniques, Sources, and Detectors, (12 August 1986); doi: 10.1117/12.936557
Show Author Affiliations
David S. Finley, University of California (United States)
Patrick Jelinsky, University of California (United States)
Stuart Bowyer, University of California (United States)
Roger F. Malina, University of California (United States)


Published in SPIE Proceedings Vol. 0689:
X-Ray Calibration: Techniques, Sources, and Detectors
Ping Lee; Paul D. Rockett, Editor(s)

© SPIE. Terms of Use