A new design - fully three dimensional -Positron Camera is presented, made of six MultiWire Proportional Chamber modules arranged to form the lateral surface of a hexagonal prism. A true coincidence rate of 56000 c/s is expected with an equal accidental rate for a 400 iCi activity uniformly distributed in a --3k water phantom. A detailed Monte Carlo program has been used to investigate the dependence of the spatial resolution on the geometrical and physical parameters. A spatial resolution of 4.8 mm FWHM has been obtained for a 18F point-like source in a 10 cm radius water phantom. The main properties of the limited angle reconstruction algorithms are described in relation to the proposed detector geometry.

Paper Details

Date Published: 1 November 1982
PDF: 8 pages
Proc. SPIE 0375, Medical Imaging and Image Interpretation, (1 November 1982); doi: 10.1117/12.934606
Show Author Affiliations
Victor Perez-Mendez, Lawrence Berkeley Laboratory and UCSF (United States)
Alberto Del Guerra, Lawrence Berkeley Laboratory (United States)
Walter R. Nelson, Stanford Linear Accelerator Center (United States)
Kwok C. Tam, General Electric Co. (United States)