Share Email Print
cover

Proceedings Paper

Comparison Of Digital Subtraction Video Densitometry And Area Length Method In The Determination Of Left Ventricular Ejection Fraction
Author(s): O. Nalcioglu; J. A. Seibert; W. W. Roeck; W. L. Henry; J. M. Tobis; W. D. Johnston
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

Digital subtraction video densitometry is utilized in the determination of left ventricular ejection fraction. The mask mode subtracted video images obtained after a peripheral intravenous injection of a bolus of Renografin 76 are used in the derivation of a density time curve. The left ventricular ejection fraction (LVEF) is computed by an algorithm described in this paper. Results for 15 patients are compared to the LVEF's obtained by applying the area-length method to the same patient data. The left ventricular ejection fractions computed by both methods had a good correlation (r=0.96) in fifteen patients.*

Paper Details

Date Published: 4 November 1981
PDF: 5 pages
Proc. SPIE 0314, Digital Radiography, (4 November 1981); doi: 10.1117/12.933061
Show Author Affiliations
O. Nalcioglu, University of California (United States)
J. A. Seibert, University of California (United States)
W. W. Roeck, University of California (United States)
W. L. Henry, University of California (United States)
J. M. Tobis, University of California (United States)
W. D. Johnston, University of California (United States)