Digital Imaging For Radiation Therapy Verification
Author(s): S. C. Taborsky; W. C. Lam; R. E. Sterner; G. M. Skarda
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

A linear diode array was used to scan a radiation field for pulsed high energy x-ray imaging. With a signal of about 50 pico-coulombs of charge from each diode for each beam pulse, the signal to noise ratio for an image of an open field was 119. An image of a porcine femur in a wax phantom was obtained by scanning the transmitted radiation. The resultant electronic image was superior in quality when compared to the standard radiotherapy port film. The simplicity of the detector system has demonstrated the feasibility of constructing an electronic portal imaging system for clinical use.

Paper Details

Date Published: 4 November 1981
PDF: 8 pages
Proc. SPIE 0314, Digital Radiography, (4 November 1981); doi: 10.1117/12.933040