Design And Physical Characteristics Of A Digital Chest Unit
Author(s): R. A. Mattson; R. A. Sones; J. B. Stickney; M. M. Tesic; G. T. Barnes
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

A prototype digital chest unit capable of imaging patients in the conventional upright position is described. The unit employs a scanning slit configuration and discrete detector array. The general features of the detector and data acquisition system are discussed as well as the unit's physical performance parameters and geometry. The resolution along and transverse to the scanning direction is in agreement with theoretical predictions. Measurements of the digital chest unit's signal-to-noise ratio versus detector exposure, detector and data acquisition dynamic range, scatter transmission and exposure linearity are also discussed.

Paper Details

Date Published: 4 November 1981
PDF: 4 pages
Proc. SPIE 0314, Digital Radiography, (4 November 1981); doi: 10.1117/12.933039