Determination Of X-Ray Tube Potential (kV) Waveform By A Noninvasive Evaluation Of Radiation Output (NERO)
Author(s): William E. Simon; Doug Richards
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

A method has been developed to noninvasively measure the instantaneous potential (kV) applied to an x-ray tube. The method uses differentially filtered x-ray detectors whose outputs during the exposure are converted to digital signals and stored in a memory array. The conversions are made every 125 microseconds. After the exposure, a microprocessor calculates the ratios of the detector outputs; computes the kV waveform from stored calibration data; and digitally displays the kV value. A resolution of ±0.5 kV at 110 kV has been achieved.

Paper Details

Date Published: 16 July 1981
PDF: 4 pages
Proc. SPIE 0273, Application of Optical Instrumentation in Medicine IX, (16 July 1981); doi: 10.1117/12.931797