Share Email Print
cover

Proceedings Paper

A non-invasive blood glucose meter design using multi-type sensors
Author(s): D. Nguyen; Hienvu Nguyen; Janet Roveda
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

In this paper, we present a design of a multi optical modalities blood glucose monitor. The Monte Carlo tissues optics simulation with typical human skin model suggests the SNR ratio for a detector sensor is 104 with high sensitivity that can detect low blood sugar limit at 1 mMole/dL ( <20 mg/dL). A Bayesian filtering algorithm is proposed for multisensor fusion to identify whether e user has the danger of having diabetes. The new design has real time response (on the average of 2 minutes) and provides great potential to perform real time monitoring for blood glucose.

Paper Details

Date Published: 17 October 2012
PDF: 6 pages
Proc. SPIE 8508, Medical Applications of Radiation Detectors II, 85080E (17 October 2012); doi: 10.1117/12.929108
Show Author Affiliations
D. Nguyen, MIAO IC Design LLC (United States)
Hienvu Nguyen, MIAO IC Design LLC (United States)
Janet Roveda, MIAO IC Design LLC (United States)
The Univ. of Arizona (United States)


Published in SPIE Proceedings Vol. 8508:
Medical Applications of Radiation Detectors II
H. Bradford Barber; Hans Roehrig, Editor(s)

© SPIE. Terms of Use
Back to Top