Evaluation of automatic exposure control performance in full-field digital mammography systems using contrast-detail analysis
Author(s): Ivan M. Suarez Castellanos; Richard Kaczmarek; Claudia C. Brunner; Hugo de Las Heras; Haimo Liu; Kish Chakrabarti
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Full Field Digital Mammography (FFDM) is increasingly replacing screen-film systems for screening and diagnosis of breast abnormalities. All FFDM systems are equipped with an Automatic Exposure Control (AEC) which automatically selects technique factors to optimize dose and image quality. It is therefore crucial that AEC performance is properly adjusted and optimized to different breast thicknesses. In this work, we studied the AEC performance of three widely used FFDM systems using the CDMAM and QUART mam/digi phantoms. We used the CDMAM phantom to generate Contrast-Detail (C-D) curves for each AEC mode available in the FFDM systems under study for phantoms with equivalent X-Ray attenuation properties as 3.2 cm, 6 cm and 7.5 cm thick breasts. Generated C-D curves were compared with ideal C-D curves constructed using a metric referred to as the k-factor which is the product of the thickness and the diameter of the smallest correctly identified disks in the CDMAM phantom. Previous observer studies have indicated that k-factor values of 60 to 80 μm2 are particularly useful in demonstrating the threshold for object detectability for detectors used in digital mammography systems. The QUART mam/digi phantom was used to calculate contrast-to-noise ratio (CNR) values at different phantom thicknesses. The results of the C-D analysis and CNR measurements were used to determine limiting CNR values intended to provide a threshold for proper image quality assessment. The results of the Contrast-Detail analysis show that for two of the three evaluated FFDM systems, at higher phantom thicknesses, low contrast signal detectability gets worse. This agrees with the results obtained with the QUART mam/digi phantom, where CNR decreases below determined limiting CNR values.

Paper Details

Date Published: 3 March 2012
PDF: 13 pages
Proc. SPIE 8313, Medical Imaging 2012: Physics of Medical Imaging, 83134I (3 March 2012); doi: 10.1117/12.912167
Show Author Affiliations
Ivan M. Suarez Castellanos, The George Washington Univ. (United States)
U.S. Food and Drug Administration (United States)
Richard Kaczmarek, U.S. Food and Drug Administration (United States)
Claudia C. Brunner, U.S. Food and Drug Administration (United States)
Hugo de Las Heras, U.S. Food and Drug Administration (United States)
QUART GmbH (Germany)
Haimo Liu, U.S. Food and Drug Administration (United States)
Kish Chakrabarti, U.S. Food and Drug Administration (United States)


Published in SPIE Proceedings Vol. 8313:
Medical Imaging 2012: Physics of Medical Imaging
Norbert J. Pelc; Robert M. Nishikawa; Bruce R. Whiting, Editor(s)

© SPIE. Terms of Use
Back to Top