Signal-to-noise ratio scaled coherence weighting for photoacoustic array imaging
Author(s): Yu-Hsin Wang; Pai-Chi Li
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Ultrasound array systems with conventional delay-and-sum beamformers are often adopted for high frame rate photoacoustic imaging. In this report, we propose a coherence factor (CF) based imaging approach to further improve the image contrast of such array systems by suppressing sidelobes of the acoustic diffraction pattern. Specifically, minimum variance based coherence factor (CFMVDR) is used. Furthermore, because CF-based weighting is susceptible to variations in signal-to-noise-ratio (SNR), we also adopt a Wiener filter approach to alleviate this problem so that the method can perform well under all SNR conditions. This is of particular interest as the SNR in photoacoustic imaging is typically low. To test this method, a human hair and a graphite phantom were used as test subjects. The imaging system consisted of a 523nm pulsed laser, and a 128-channel linear array (bandwidth from 5 to 10MHz) for photoacoustic signal detection. It is demonstrated that the beam widths (i.e., lateral resolution) can be effectively improved and the noise background is suppressed by 20 dB. The contrast improvement is also evident.

Paper Details

Date Published: 23 February 2012
PDF: 6 pages
Proc. SPIE 8223, Photons Plus Ultrasound: Imaging and Sensing 2012, 822340 (23 February 2012); doi: 10.1117/12.909854
Show Author Affiliations
Yu-Hsin Wang, National Taiwan Univ. (Taiwan)
Pai-Chi Li, National Taiwan Univ. (Taiwan)


Published in SPIE Proceedings Vol. 8223:
Photons Plus Ultrasound: Imaging and Sensing 2012
Alexander A. Oraevsky; Lihong V. Wang, Editor(s)