The quasi-static method with application of half-sine pressure pulse is presented to calibrate the piezoelectric sensor, which is used for the dynamic pressure measurement of weapons. A pressure generator based on the drop hammer hydraulic system is manufactured to get the half-sine pressure pulse. The oil cylinder of the generator is reconstructed to install four standard pressure sensors and two calibrated sensors simultaneously. With pressure taken from four standard sensors as calibrating excitation, and response data obtained from calibrated sensors, the working sensitivities of sensors are worked out through regression analysis. The experimental results obtained with sensor 6215 at the national shooting range shows that it is effective to calibrate piezoelectric sensors using half-sine pressure pulse. The residual standard deviation of the equation fitting is less than 0.7%; the linearity is less than 0.21%; and the relative uncertainty of the four standard sensors is less than 0.7%, under the precision target of the calibration system acceptance.

Paper Details

Date Published: 29 December 2010
PDF: 6 pages
Proc. SPIE 7544, Sixth International Symposium on Precision Engineering Measurements and Instrumentation, 75445Y (29 December 2010); doi: 10.1117/12.885491
Show Author Affiliations
Qicheng Zhang, Nanjing Univ. of Science and Technology (China)
Nanjing Institute of Technology (China)
Fei Shang, Nanjing Univ. of Science and Technology (China)
Deren Kong, Nanjing Univ. of Science and Technology (China)


Published in SPIE Proceedings Vol. 7544:
Sixth International Symposium on Precision Engineering Measurements and Instrumentation
Jiubin Tan; Xianfang Wen, Editor(s)