In order to improve the accuracy of an instrument by reducing or correcting the error of a sensor system, a calibration device which can be used to trace the laser wavelength is used to calibrate the static and dynamic characteristics of dual sensor system. Based on the accurate calibration data of the sensor, the nonlinearity error and the coherence error are corrected and compensated using least-square approximation model and interpolation method. Test results indicate that the nonlinearity of each sensor can be within 0.15% in the range of ±10μm and the incoherence is less than 0.2 % in the range of ±5μm after corrected. Meanwhile, the cause and effect of a misalignment error between dual sensors are analyzed.

Paper Details

Date Published: 31 December 2010
PDF: 6 pages
Proc. SPIE 7544, Sixth International Symposium on Precision Engineering Measurements and Instrumentation, 75445H (31 December 2010); doi: 10.1117/12.885462
Show Author Affiliations
Yong-bo Wu, China Jiliang Univ. (China)
Zi Xue, National Institute of Metrology (China)
Wei Zhan, National Institute of Metrology (China)