Share Email Print
cover

Proceedings Paper

High-performance HPD for photon counting
Author(s): A. Fukasawa; A. Kamiya; S. Muramatsu; Y. Negi; M. Suyama
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

We have developed a new Hybrid Photo-Detector (HPD) with relatively large effective area, fast time response, high timing resolution, high quantum efficiency and extremely low probability of afterpulse. A GaAsP (Gallium Arsenide Phosphide) photocathode type has high quantum efficiency approximately 45% around the wavelength of 500 nm, and the size of the effective area is 3 mm in diameter. The pulse height for single photon is approximately +2 mV with 50 ohm load impedance, and it can be easily observed with a fast response oscilloscope. The rise and fall times for impulse light are approximately 0.4 ns. The timing resolution for single photon for the GaAsP photocathode type was estimated to be approximately 80 ps.

Paper Details

Date Published: 13 May 2011
PDF: 8 pages
Proc. SPIE 8033, Advanced Photon Counting Techniques V, 80330S (13 May 2011); doi: 10.1117/12.883605
Show Author Affiliations
A. Fukasawa, Hamamatsu Photonics K.K. (Japan)
A. Kamiya, Hamamatsu Photonics K.K. (Japan)
S. Muramatsu, Hamamatsu Photonics K.K. (Japan)
Y. Negi, Hamamatsu Photonics K.K. (Japan)
M. Suyama, Hamamatsu Photonics K.K. (Japan)


Published in SPIE Proceedings Vol. 8033:
Advanced Photon Counting Techniques V
Mark A. Itzler; Joe C. Campbell, Editor(s)

© SPIE. Terms of Use