CZT detector in multienergy x-ray imaging with different pixel sizes and pitches: Monte Carlo simulation studies
Author(s): Yu-Na Choi; Hee-Joung Kim; Hyo-Min Cho; Chang-Lae Lee; Hye-Suk Park; Dae-Hong Kim; Seung-Wan Lee; Hyun-Ju Ryu
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

A photon counting detector based on semiconductor materials is a very promising approach for x-ray imaging. Cadmium zinc telluride (CZT) semiconductor has a high atomic number which results in higher absorption coefficients for x-rays. However, the CZT detectors exhibit several problems with hole trapping and charge sharing. Charge sharing occurs due to diffusion of charge and characteristic x-ray escape and scattered x-rays in the detectors. In this study, we evaluated the effect of interaction with CZT detector using Monte Carlo simulations. To demonstrate the effectiveness of CZT detector in clinical application, we reported confirmation of CNR improvement in K-edge images, and material decomposition using energy selective windows. X-ray energy spectrum acquired at 120 kVp tube voltage and 2 mm Al filtration and 10 cm added water phantom in the x-ray beam. Geant4 Application for Tomographic Emission (GATE) version 6.0 was used for a CZT crystal with size of 10x10 mm2 and thickness of 4 mm. The detector pixel with sizes of 0.09x0.09, 0.45x0.45, and 0.90x0.90 mm2 were simulated. For all pixel sizes, the x-ray spectra of the simulations were distorted towards the lower energy region. Because the characteristic x-rays add counts in the range of 20-40 keV. The magnitude of this deterioration is substantial for small pixel sizes. However, we demonstrated that the distortion of spectrum does not greatly affect the x-ray imaging. The GATE simulation model and these results may be used as a basis of development of energy-resolved photon counting x-ray detector. We believe that the CZT detector may enhance the detectability of multi-energy x-ray imaging.

Paper Details

Date Published: 16 March 2011
PDF: 12 pages
Proc. SPIE 7961, Medical Imaging 2011: Physics of Medical Imaging, 79614H (16 March 2011); doi: 10.1117/12.878357
Show Author Affiliations
Yu-Na Choi, Yonsei Univ. (Korea, Republic of)
Hee-Joung Kim, Yonsei Univ. (Korea, Republic of)
Hyo-Min Cho, Yonsei Univ. (Korea, Republic of)
Chang-Lae Lee, Yonsei Univ. (Korea, Republic of)
Hye-Suk Park, Yonsei Univ. (Korea, Republic of)
Dae-Hong Kim, Yonsei Univ. (Korea, Republic of)
Seung-Wan Lee, Yonsei Univ. (Korea, Republic of)
Hyun-Ju Ryu, Yonsei Univ. (Korea, Republic of)


Published in SPIE Proceedings Vol. 7961:
Medical Imaging 2011: Physics of Medical Imaging
Norbert J. Pelc; Ehsan Samei; Robert M. Nishikawa, Editor(s)

© SPIE. Terms of Use
Back to Top