Study on peak-seeking algorithm for wavelength in Fourier transform infrared spectrometer
Author(s): Guojin Feng; Yu Wang; Ping Li; Chundi Zheng; Qiaoxiang Zhang; Yingwei He
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

In the progress of wavenumber traceability using polystyrene films, different peak-seeking algorithms are discussed in this paper. Conclusion shows that in commonly used methods, spline method, center of gravity method and the second derivative method, spline method is better than the others with simple algorithm and good agreement. It is strongly recommended to use in practical for calibrating FTIR spectrometer.

Paper Details

Date Published: 4 November 2010
PDF: 5 pages
Proc. SPIE 7854, Infrared, Millimeter Wave, and Terahertz Technologies, 78542D (4 November 2010); doi: 10.1117/12.868618
Show Author Affiliations
Guojin Feng, National Institute of Metrology (China)
Yu Wang, National Institute of Metrology (China)
Ping Li, National Institute of Metrology (China)
Chundi Zheng, National Institute of Metrology (China)
Qiaoxiang Zhang, National Institute of Metrology (China)
Yingwei He, Beijing Institute of Technology (China)


Published in SPIE Proceedings Vol. 7854:
Infrared, Millimeter Wave, and Terahertz Technologies
Cunlin Zhang; Xi-Cheng Zhang; Peter H. Siegel; Li He; Sheng-Cai Shi, Editor(s)