Quantifying breast density with a cone-beam breast CT
Author(s): Xinhua Li; Bob Liu
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Volumetric breast density was evaluated using a simulated cone beam breast CT with 80 kVp. The breast was modeled as a cylinder with background tissue composition of 20% glandular and 80% adipose. Various objects with different sizes and tissue compositions were embedded. Ray-tracing algorithm was utilized to obtain projection images in a full rotation without considering scatter, beam hardening and imaging noise. Filtered backprojection was adopted for image reconstruction with high quality. Reconstructed images had flat profiles except at large cone angle of 8.6° to 10°. They were calibrated using known linear attenuation coefficients of two image contrast objects. A 3D mapping of tissue densities could be directly computed within 5% error. Tissue volumes were obtained by counting voxels in appropriate attenuation coefficient ranges. Results of contrast objects were consistent with true volumes within 10% error. However, cone angle artifact decreased pixel values, and a reduction algorithm was required for accurate tissue assessment at large cone angles. This study indicates the possibility of excellent quantitative breast density measurements and volume assessments with cone-beam breast CT.

Paper Details

Date Published: 23 March 2010
PDF: 11 pages
Proc. SPIE 7622, Medical Imaging 2010: Physics of Medical Imaging, 762245 (23 March 2010); doi: 10.1117/12.844369
Show Author Affiliations
Xinhua Li, Massachusetts General Hospital (United States)
Bob Liu, Massachusetts General Hospital (United States)


Published in SPIE Proceedings Vol. 7622:
Medical Imaging 2010: Physics of Medical Imaging
Ehsan Samei; Norbert J. Pelc, Editor(s)

© SPIE. Terms of Use
Back to Top