Toward an international consensus strategy for periodic quality control of digital breast tomosynthesis systems
Author(s): J. Jacobs; N. Marshall; L. Cockmartin; F. Zanca; R. van Engen; K. Young; H. Bosmans; E. Samei
Format Member Price Non-Member Price
PDF $17.00 $21.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

As a collaborative effort between scientists affiliated with the American Association of Physicists in Medicine (AAPM) and the European reference center for breast cancer screening and diagnosis (EUREF), the Working Group on Phantoms for Breast Imaging (WGPBI) aims to develop phantoms and evaluation techniques for 2D & 3D breast imaging modalities. In the first phase of this collaboration, this project aimed to develop a phantom and associated procedure for constancy testing of digital breast tomosynthesis (DBT) systems. The procedure involves daily and weekly components. The daily evaluation is performed on a simple, homogenous PMMA plate of 4 cm thickness. For the weekly part, a new phantom has been designed consisting of a 45 mm thick homogeneous slab of PMMA with a set of spherical and rectangular inserts at specific 3D positions, and a thin wire positioned at a small angle to the plane of the detector. Quality control parameters are extracted from both projection images (if available) and reconstructed planes. The homogeneous phantom for daily QC allows a trend analysis of homogeneity and the assessment of detector artifacts. With the proposed phantom concept for weekly QC, the stability of the following parameters can be evaluated: the propagation and correlation of the noise in plane and across the reconstructed tomographic planes, lag, signal difference to noise ratio (SDNR) and signal to noise ratio (SNR), the geometry and the motion, effective thickness of the reconstructed planes, homogeneity, distance accuracy, frequency dependent SNR, and artifacts. Analysis of the DICOM header provides information on the stability of the automatic exposure control (AEC), exposure settings, and several system parameters. In an on-going study, the proposed strategy is being applied to five tomosynthesis systems both in Europe and in the US. In this paper we report on the specifics of the phantom, the QC procedure, the practicalities of remote data analysis, and the results of the initial trial.

Paper Details

Date Published: 18 March 2010
PDF: 11 pages
Proc. SPIE 7622, Medical Imaging 2010: Physics of Medical Imaging, 76220G (18 March 2010); doi: 10.1117/12.844228
Show Author Affiliations
J. Jacobs, Univ. Hospitals Leuven (Belgium)
N. Marshall, Univ. Hospitals Leuven (Belgium)
L. Cockmartin, Univ. Hospitals Leuven (Belgium)
F. Zanca, Univ. Hospitals Leuven (Belgium)
R. van Engen, Radboud Univ. Nijmegen Medical Ctr. (Netherlands)
K. Young, The Royal Surrey County Hospital (United Kingdom)
H. Bosmans, Univ. Hospitals Leuven (Belgium)
E. Samei, Duke Univ. (United States)


Published in SPIE Proceedings Vol. 7622:
Medical Imaging 2010: Physics of Medical Imaging
Ehsan Samei; Norbert J. Pelc, Editor(s)

© SPIE. Terms of Use